16
TROUBLESHOOTING
Problem
Possible Cause
Corrective Action
You smell or taste contaminants
or an irritation occurs.
Misuse, improper assembly or
malfunction of equipment.
Leave work area immediately
and contact your supervisor.
Do not use the PAPR until the
cause is identified and corrected.
Blower does not run when switch
is depressed
Battery is discharged
Faulty power switch
Faulty motor
Recharge 14-16 hours
Replace battery pack
Replace motor/blower
Battery does not accept charge,
but PAPR unit runs when
plugged into charger.
Faulty battery connection or
terminals dirty/damaged.
Check that battery terminals are
clean and battery is properly
attached and locked into the
PAPR housing.
PAPR fails airflow test
Clogged filter
Battery needs charging
Battery does not hold charge
Breathing tube blocked
Replace filter
Charge battery
Replace battery
Locate and remove restriction