Lumencor SOLA U-NIR is switched off.
•
Switch light source on, otherwise contact 3DHISTECH Service for support.
9.5
Maintenance
This chapter describes the most common maintenance solutions that can be performed by the
user of the product. For maintenance solutions not described in this manual, contact the service
personnel.
9.5.1
Removing the protective covers
The protective covers of the Pannoramic® SCAN RX must be opened to perform any
maintenance work, and also to eliminate errors, for example slide or magazine jamming.
Before removing the covers, exit Pannoramic® Scanner Software for Research control
software, switch off the 24 V power supply and disconnect the power cables.
Warning!
Avoid leaning over the device when opening it.
Figure 130. Removal of covers
Five eccentric screws
(3)
are located on the bottom face of the base plate
(4)
. They attach
the covers to the base plate.