PANNORAMIC® DESK DW II 3.0.4
TROUBLESHOOTING AND MAINTENANCE
MARCH 30, 2021
99
8
Troubleshooting and Maintenance
8.1
Recovery functions
PSW automated backups allow you to continue scanning after unexpected events within as
little time as possible. After the program is restarted and the microscope is initialized, you
can confirm the automatic restore of pre-event state by clicking
Yes
.
Figure 82 – Recovery confirmation window
Warning!
Click
Yes
only if you are sure that everything is completely unchanged after the previous run.
If anything has changed or you are unsure, click
No
.
For more comfortable and safer use, the
Planner autosave
function allows the taking of
automatic backups while creating a scan plan. The backups are placed in the default
document folder of the actual windows user and can be loaded by clicking the
Load plan
button.
8.2
Safety information
Check cables, plugs and connections before operation, and if defective or cannot be
replaced, contact 3DHISTECH Service for support.
Check whether the device is compatible with your local line voltage. It can operate at line
voltages 100V to 240V and 50/60 Hz.
Only plug the device in a socket with protective grounding (earth) contact.
Do not operate the device in explosive environment.
Disconnect the device from the power outlet before opening it or replacing a part.