iSaCS – IHC Stainer, Automated Coverslipper and Scanner
User's Guide
4 Commissioning
4.5.5 Filling up the reagent chain with pipette tips and reagent vials
1. Open the upper door
2. Remove used pipette tips from the racks at right side of the chain (
1
), then insert unused
ones into the racks at the left side of the chain (
2
).
Figure
10 – Inserting pipette tips and reagent vial
3. With their caps removed, insert reagent vials into their racks of the chain (
3
)
4.5.6 Filling the coverslipper unit
1. Open the upper door
2. Replace the used pipette tip with a new one (
1
)
Note:
The unused pipette tip must be placed to the lower collar of the pipette
rack.
3. Insert the vial of the mounting material to its place (
2
)
4. Fill up the coverslip container unit with the necessary amount (
3
)
Figure 11 – Locations of pipette tips and the reagent vial
September 22, 2016
3DHISTECH Ltd.
25(51)