Pannoramic DESK II 1.22 User’s Guide

5 Troubleshooting and Maintenance

5. Reattach protective covers.

6. Connect power plug and turn power supply on.

7. Launch the scanner software.

Black image – Illumination error! Probably the lamp has burned out. Please turn off the scanner 
and check the lamp!

Halogen lamp is defective

Replace the halogen lamp (see section 

5.3.2 Replacing the 

 

 halogen lamp

).

Virtual slide is improperly focused almost across its entire surface

Focus range enabled

Disable 

Limiting focus range

.

Poor image quality

Imaging error (for example, chromatic aberrations)

Contact 3DHISTECH Service for support.

  Warning!

Do not modify the camera driver settings, as the default factory 
settings give the most accurate images.

Virtual slides show poor stitching in Pannoramic Viewer

Scan camera not properly adjusted (for example, due to loose fixture)

Contact 3DHISTECH Service for support.

June 24, 2016 - Rev. 1

3DHISTECH Ltd.

66

(

76

)

Summary of Contents for Pannoramic DESK II

Page 1: ...PDII122 UG Rev1 Pannoramic DESK II 1 22 User s Guide June 24 2016 Rev 1 ...

Page 2: ...tions 16 1 2 3 Control Software 18 2 Installation 19 2 1 Control computer ports 19 2 2 Connectors and cables 20 3 Preparing Slides 21 3 1 Affixing Barcode Stickers to Slides 22 3 2 Inserting and Removing Slides 23 4 Software Description 26 4 1 Starting Exiting Pannoramic DESK II Software Application 26 4 1 1 Starting the software 26 4 1 2 Exiting the software 26 4 2 Options 26 4 2 1 Barcode Settin...

Page 3: ...o Server 58 4 5 Using Profiles 63 5 Troubleshooting and Maintenance 64 5 1 Safety Information 64 5 1 1 Cleaning the device 64 5 1 2 Preventive maintenance 65 5 2 Troubleshooting 65 5 3 Maintenance 68 5 3 1 Removing the protective cover 68 5 3 2 Replacing the halogen lamp 69 5 3 3 Removing a slide 71 5 3 4 Transporting Pannoramic DESK II 72 6 Technical Data 74 Index 78 June 24 2016 Rev 1 3DHISTECH ...

Page 4: ... IP are the sole properties of the 3DHISTECH Ltd 3DHISTECH Ltd is not liable for damage of whatever nature including but not limited to general or specific damage indirect damage consequential damage or incidental damage including the results of the analysis of the digitized slides for example change of health status related to erroneous diagnosis from the digitized slide s that stems from or is a...

Page 5: ... KIND EXPRESSED IMPLIED OR OTHERWISE INCLUDING WITHOUT LIMITATION ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN NO EVENT SHALL SAM LEFFLER OR SILICON GRAPHICS BE LIABLE FOR ANY SPECIAL INCIDENTAL INDIRECT OR CONSEQUENTIAL DAMAGES OF ANY KIND OR FOR ANY DAMAGES WHATSOEVER RESULTING FROM LOSS OF USE DATA OR PROFITS WHETHER OR NOT ADVISED OF THE POSSIBILITY OF DAMAGE AND ON A...

Page 6: ...101 Particular requirements for in vitro diagnostic IVD medical equipment EN 55011 1998 Group 1 Class A Limits and methods of measurement of radio disturbance characteristics of industrial scientific and medical ISM radio frequency equipment This product has been tested to the requirements of CAN CSA C22 2 No 61010 1 second edition including Amendment 1 or a later version of the same standard inco...

Page 7: ...uch as F11 or the Ctrl M key combination Example Cross references within this document or to other documents Warning It might cause irreversible damage or harm to the user of the product or the product itself if this instruction is not followed Caution It might cause the product to not work properly if this instruction is not followed and alerts to a potential danger situation Note Calls attention...

Page 8: ... All claims for warranty and warranty services will be null and void in any such case Warning Insert the power cord plugs only into a grounding outlet The protective action must not be rendered ineffective by extension cables without protective conductor The Pannoramic DESK II has a built in protection which must not be rendered ineffective by extension cables without protective earth ground condu...

Page 9: ...over on Caution Avoid touching the lamp bulb Disconnect power plug and let the lamp cool down for about 15 minutes before replacing the lamp Crush hazard Never open the front door unless prompted accordingly by software When you open the door at any other time make sure not to reach with your hands into the instrument to avoid the risk of crushing your fingers Warning The front door is equipped wi...

Page 10: ... accepted in cases of natural wear notably wear and tear parts or improper product handling The Product Manufacturer will refuse any liability in the event of damage resulting from faulty operation negligence or unauthorized intervention into the product notably from the replacement of product parts or the usage of accessory parts from other manufacturers Any such case will void the user s claim f...

Page 11: ... Virtual Slide Slide A digital image of a thin glass plate on which specimens are mounted for microscopic study a dynamic interactive image that you can manage save magnify zoom name evaluate annotate mark and comment send to a colleague electronically for co operation or advice and so on June 24 2016 Rev 1 3DHISTECH Ltd 11 76 ...

Page 12: ... the functionality of the scanner program 1 1 2 Features and Benefits The Pannoramic DESK II offers world class brightfield whole slide scanning An outstanding 0 12 µm pixel resolution is achieved with the 40 0 95 NA equivalent to 80 magnification Carl Zeiss Plan Apochromat objective Key features of the Pannoramic DESK II Scanner are Capable of scanning large slides 3MP camera for barcode reading ...

Page 13: ...escription 1 1 3 Warning and Information Labels The following figure contains the labels used on the Pannoramic DESK II main unit and the external mains power supply unit Figure 1 Warning and information labels June 24 2016 Rev 1 3DHISTECH Ltd 13 76 ...

Page 14: ... s Guide 1 Product Description 1 1 4 Base Unit The following figure shows the main parts of the base unit Figure 2 Main parts of the base unit 1 Protective cover 2 Sliding door 3 Scan camera June 24 2016 Rev 1 3DHISTECH Ltd 14 76 ...

Page 15: ...lowing hardware and software items 1 2 1 Pannoramic DESK II system The following figure shows the main components of the Pannoramic DESK II system Figure 3 Pannoramic DESK II system overview 1 24 V external power supply unit 2 Pannoramic DESK II base unit 3 Monitor 4 Mouse 5 Control computer 6 Keyboard June 24 2016 Rev 1 3DHISTECH Ltd 15 76 ...

Page 16: ... Point Grey Grasshopper 3 Color 1 CCD monochrome camera for normal brightfield scanning Adimec QUARTZ Q 12A180 1 CMOS camera for high quality brightfield applications allowing fast scanning Camera adapters 0 63 adapter for Point Grey Grasshopper 3 camera only 1 adapter for Point Grey Grasshopper 3 camera only 1 6 adapter for Adimec camera only June 24 2016 Rev 1 3DHISTECH Ltd 16 76 ...

Page 17: ...thorized person Base Unit options Pannoramic DESK II Base Brightfield Option base unit equipped with Point Grey Grasshopper 3 color camera for Brightfield scanning Pannoramic DESK II Fast Brightfield Option base unit equipped with Adimec QUARTZ Q 12A180 camera for Brightfield scanning Objectives Carl Zeiss Plan Apochromat 20 NA 0 8 Carl Zeiss Plan Apochromat 40 Corr NA 0 95 Warning Do not change c...

Page 18: ...nal Extended focus Z stack Barcode reader licensed feature Pannoramic Viewer software fully functional Optional CaseCenter Pannoramic Viewer TMA module 3D Reconstruction module QuantCenter HistoQuant NuclearQuant MembraneQuant and DensitoQuant modules TumorBoard module E School Encoder June 24 2016 Rev 1 3DHISTECH Ltd 18 76 ...

Page 19: ... an installation carried out by an unauthorized person Only qualified service assistants are allowed to install the hardware and software The product is ready for use after installation 2 1 Control computer ports The following figure shows the location of the connectors in order to enable easier transportation of the product for a short distance by disconnecting cables and detaching hardware compo...

Page 20: ...control computer Preview camera USB cable between scanner and control computer Adimec CoaXPress cables between Adimec camera and control computer Point Grey Grasshopper 3 Color USB 3 0 cable between Point Goint Grey Grasshopper 3 camera and control computer Video cable between monitor and control computer Power supply cable between 230V 110V and monitor Power supply cable between 230V 110V and con...

Page 21: ...slide when it is broken or its edges are not properly cut or grounded Using such slides can lead to injury Warning Ensure that no part of the mounted coverslip protrudes from the edges of the slide Use slides with completely dried embedding medium only Otherwise the coverslip can be shifted that makes slide insertion less reliable Not completely dried embedding might also influence the operation o...

Page 22: ...e barcode sticker 2 from the roll 2 Stick it on the colored label area 1 of the slide on the specimen side Caution Ensure that a marginal space of 1 2 mm remains on all sides between the sticker and the label area outer limits It is vital for barcode identification Do not let barcode sticker protrude from slide edges or stick on the coverslip surface 3 The thickness of the barcode sticker on the s...

Page 23: ...e tray The edge of the broken area of the slide is sharp and its removal can lead to injury and contagion Broken slide can be removed only by a well trained person Figure 6 Inserting and removing a slide 1 Click in Pannoramic DESK II control software The specimen holder moves to the load unload position 2 Open the slide insert door 1 by pushing it to the left Caution When opening the slide insert ...

Page 24: ...er Incorrect slide insertion may break the glass The edge of the broken area of the slide is sharp and its removal can lead to injury and contagion Broken slide can be removed only by a well trained person 5 Close the fixing lever 2 by pushing it to the right so that it touches and therefore fixes the slide 6 Close the slide insert door 1 by pushing it to the right Caution When closing the slide i...

Page 25: ...SK II control software The specimen holder moves to the load unload position 2 Open the slide insert door 1 by pushing it to the left 3 Open the fixing lever 2 by pulling it to the left 4 Remove the slide 4 5 Insert another slide or close the fixing lever 2 and the door 1 Note The slide insert door is equipped with an interlock switch It automatically stops a running digitization process when the ...

Page 26: ...fter successful initialization the program displays a Mode selection window Figure 7 Mode selection window The selection window displays Manual Brightfield scan mode for digitizing individual slides If you had exited this mode and want to reactivate it click Manual 4 1 2 Exiting the software If you close a mode or one of the options window the Mode selection window turns active again To exit the s...

Page 27: ...Barcode settings Barcode settings can be modified in the Barcode Settings window By predefining the barcode type foreground color black or white the color of the barcode and other settings you enable the Read Barcode option when using the control software June 24 2016 Rev 1 3DHISTECH Ltd 27 76 ...

Page 28: ...t the slide is not tilted and that it rests perfectly horizontally on the stage To display an image of the barcode area click You can redefine the barcode reading area by dragging the edges of the green square To check whether the barcode reading settings are correct display the barcode digits in the Barcode field by clicking Note If you did not set the barcode type correctly or there is no barcod...

Page 29: ... Code 128 Code 16K Interleaved 2 of 5 Standard 2 of 5 IATA 2 of 5 Matrix 2 of 5 Telepen Code 93 Code 11 MSI Codabar Patch Code Pharmacode EAN 13 EAN 8 UPC A UPC E 2D Barcode Types You can select the following 2D barcode types Aztec Data Matrix PDF417 Compact Micro PDF417 QR Micro QR MaxiCode June 24 2016 Rev 1 3DHISTECH Ltd 29 76 ...

Page 30: ...ick Use barcode compensation image to use the previously created compensation image for scanning the label area The compensation image improves the effectiveness of barcode reading 1 Prepare a slide with a label area covered fully by a white sticker 2 Take a new compensation image of the blank barcode sticker by clicking 4 2 2 Image Compensation You can use a saved compensation image whenever comp...

Page 31: ...hold settings Therefore the scanner will not use those areas for the creation of the compensation images Current threshold To modify the threshold settings set the desired threshold level in the box Use Selection Box Select this if you have a compensation slide that has only certain predefined areas for creating compensation images due to the type of the compensation slide or because the compensat...

Page 32: ...Pannoramic DESK II 1 22 User s Guide 4 Software Description Image Browser panel Figure 11 Image Browser This function displays the created compensation image June 24 2016 Rev 1 3DHISTECH Ltd 32 76 ...

Page 33: ...tion about the setting you are interested in Base Settings Figure 12 Base settings Objective type You can select the type and the magnification level of the objectives Camera type Select the correct camera type in Pos 1 Camera type then the software automatically loads the camera settings upon clicking the button June 24 2016 Rev 1 3DHISTECH Ltd 33 76 ...

Page 34: ...andard Gamma value White Point D50 Legacy a Color scheme calibrated for the previous software version is applied Custom parameters are freely customizable as follows Color Scheme Calibrated Scheme is defined by the color calibration slide Legacy a Color scheme calibrated for the previous software version is applied Raw the color information of the recorded raw image is applied White Point D65 colo...

Page 35: ...SK II 1 22 User s Guide 4 Software Description Filter Settings Figure 13 Filter settings The settings on this panel apply to Fluorescent mode therefore irrelevant and inactive June 24 2016 Rev 1 3DHISTECH Ltd 35 76 ...

Page 36: ...amera rotation Caution If you have transported the product or changed the camera there is a possibility that the camera does not display the images properly If you experience such problems contact service support Do not modify these settings yourself June 24 2016 Rev 1 3DHISTECH Ltd 36 76 ...

Page 37: ...lication call settings for the following cases Slide Scanning Finished Scanning Error Occurred For Slide Scanning Finished and Scanning Error Occurred you can define the external software by entering the file path in the Application box or by browsing for it by clicking If you define parameters in the Parameters box it will pass them to the software that it starts If no parameter is defined then t...

Page 38: ...erver add or remove users and also set the message body in the Message subject field by entering a text for example to inform recipients on the location of the scanner or the slide set you are working with Click Test to test settings by sending an e mail to the address defined in the E mail To field after you have finished to check their relevancy before applying them First click Setup to display ...

Page 39: ...s button to send test e mail to the specified address There are 2 feedback messages that come up after sending the e mail An e mail has been sent to the recipient An error has been occurred The e mail could not be sent Check parameters then retry If you have properly set messenger settings click OK then click Users in the Message Call section of the Event Call Settings window to address functions ...

Page 40: ... settings You can also set profile settings here and start scanning in Brightfield Manual mode For more information on profiles see section 4 5 Using Profiles Figure 18 Main window of control software 1 Tab panel with scanner functions and settings 2 Scan camera live image display area 3 Preview image recorded by the preview camera 4 Navigation and Preview image editor toolbar 5 Information area a...

Page 41: ...and that it rests perfectly horizontally on the stage To enter the slide name To enter the slide name manually type it into the Slide Name field To use the barcode as a slide name if a compatible barcode sticker is affixed on the slide click First the software automatically detects the barcode on the slide then the barcode string will be displayed in the Slide Name field You can also find the butt...

Page 42: ... the current profile click Save To select an existing profile from the selection menu click Load To delete the current profile click Delete To access the Scan Options window click Note These settings are only applicable if you do not use profiles If you use profiles the settings stored in them are used 4 3 2 Scan Options window Figure 20 Scan options window June 24 2016 Rev 1 3DHISTECH Ltd 42 76 ...

Page 43: ...n enables the seamless stitching of digital tiles Save image in original Bayer size function is only available for products installed with Stingray camera It is not accessible for configurations installed with CIS camera Image fields will be saved with native pixel resolution If this function is disabled you will work with reduced pixel resolution These resolution differences are not really visibl...

Page 44: ...ver application see section 4 4 Scan To Server If you enable the Directory function the virtual slides will be digitized in the specified folder You can browse for the folder with the Browse button If you specify a folder that does not exist the software will create it All parameter inputs are saved for the next software session 4 3 3 Preview tab The Preview tab offers options for threshold settin...

Page 45: ...sensitive scanning but it is possible that some of the smaller specimen areas will not be digitized Scan area inside the marker If the specimen cannot be identified as required for digitization for example weak contrast it can help to encircle completely the specimen with a black marker pen You have to draw a full circle with the pen to enable the detection Marker pen detection will make specimen ...

Page 46: ...eas Warning Do NOT use Remove coverslip function on TMA slides The layout of TMA slides may mislead the detecting algorithm Fill holes in scan area If selected the holes inside the specimen will also be scanned Often these holes contain small sample objects that otherwise would not be scanned Remove specks If selected the scanner does not digitize objects smaller than the predetermined size This h...

Page 47: ...d preview mask or TMA grid Use selection box If selected only the area inside the box is scanned It is useful if you only want to scan the area that is equivalent with the size of the coverslip so external specks will not be scanned 4 3 4 Preview area scanned by preview camera The preview image shows the area that will be digitized with a red and or an orange mask If the Use selection box is enabl...

Page 48: ...will not be scanned These areas are not digitized on the basis of their position for example they are outside the selection box They are detected by the selected threshold value Marker If Scan area inside the marker is selected in the Preview tab the preview camera detects the marker pen The inside areas appear in red color and will be digitized Areas scanned in This color is only displayed during...

Page 49: ...king with the mouse TMA tool To undo your last action click maximum of 40 changes To position the specimen holder that holds the current slide click Position sample and then click on the area of interest on the preview image This will define the Live image that you can access from the Focus panel To display the original image created by the preview camera click To shrink or enlarge the preview ima...

Page 50: ...e selection settings on the following four tabs Select tab Figure 28 TMA tool Select To create a new TMA object click New To select an existing TMA object click on the object name The selected object name will highlight To delete an existing TMA object click Delete Block tab Figure 29 TMA tool Block June 24 2016 Rev 1 3DHISTECH Ltd 50 76 ...

Page 51: ...e Y µm fields or select the appropriate values with the arrows 4 Enter the average diameter of the spots in the Diameter of spots µm field or select the appropriate value with the arrows Grid tab The Grid tab displays the previously set grid settings Figure 30 TMA tool Grid Move tab The Move tab allows the rotation and moving of the TMA selection spots that are displayed on the preview image To mo...

Page 52: ...Pannoramic DESK II 1 22 User s Guide 4 Software Description Figure 31 TMA tool Move tab with preview June 24 2016 Rev 1 3DHISTECH Ltd 52 76 ...

Page 53: ...2 User s Guide 4 Software Description Label area To rotate the label area by 180 select the Rotate image by 180 degrees checkbox Figure 32 Label area 4 3 5 Focus tab Figure 33 Focus tab June 24 2016 Rev 1 3DHISTECH Ltd 53 76 ...

Page 54: ...uct as a traditional microscope Define the manual focus step size by adjusting the Step size slider Search for the sharpest focus on the current field of view automatically by clicking Multilayer mode Select Single layer option if you want to scan a single layer of the slide By choosing the Extended focus option the software selects the sharpest image from each focus level for each image field and...

Page 55: ...tab This tab contains control and maintenance functions Show images during scanning If you select this option you will be able to see the currently scanned image field during scanning in the live image display field This slows down digitization process and should be activated only with purpose June 24 2016 Rev 1 3DHISTECH Ltd 55 76 ...

Page 56: ... scan camera The live image is displayed if no current digitization process is running or Show images during scanning option is enabled Figure 36 Display field and toolbar for live images Navigation on the live image Click a desired area of the slide in the Live image to display that area in the center of the Live image field June 24 2016 Rev 1 3DHISTECH Ltd 56 76 ...

Page 57: ...els were detected Calculates white balance only for color cameras Saves the current image as BMP or JPEG file Fast live mode live images are displayed with a frequency of at least 10 images per second If an exposure time is set to more than 100 ms the pixel values are multiplied accordingly for example if an exposure time of 200 ms is set the pixel values are multiplied by 2 and the exposure time ...

Page 58: ...ast line Available disk space of the information area of Scan Progress You can interrupt a running digitization process by clicking This button is only enabled if a digitization process is in progress 4 4 Scan To Server The ScanToServer function is available for you to scan directly to a dedicated CaseCenter server To setup connection with a designated server follow the steps below 1 Click Scan Op...

Page 59: ...then click Connect Figure 38 Scan to CaseCenter option 3 Type in or select the server address from the Connect to drop down list and give the User name and Password Activate the Secure connection option to enable HTTPS connection Click Connect when finished Figure 39 Server Log in window June 24 2016 Rev 1 3DHISTECH Ltd 59 76 ...

Page 60: ... the Scan Options window Slide settings in the Scan To Server window 1 In the Routine work tab panel click the button 2 The Scan To Server window opens Figure 41 Scan To Server window Slide data is displayed in a form Upon activating the Read barcode if no name is defined on the list option the slide barcode will be read and set in the Barcode field If you do not want to use this option you can ty...

Page 61: ... of the barcode will be displayed and set as slide name 123 Case and Block fields can be set this way as well if necessary NOTE When filling slide data fields press TAB to move to the next cell Figure 42 Profile selection in the Scan To Server window Profile can be added by selecting a list entry from the Profile drop down list Select a folder where the slide will be scanned to After clicking the ...

Page 62: ...ically selected as scanning destination based on the CaseID containing slide and patient data stored on HIS LIS Therefore Slide name Folder Case and Block field data cannot be modified only a profile can be selected Figure 44 Scan To Server window during Barcode parsing Click Done when finished or Cancel to quit the Scan To Server window If you save the form slides will be scanned to server accord...

Page 63: ...al Bayer size ON OFF available if Point Grey Grasshopper 3 camera is used Determination of the area to be scanned Use of automatic threshold specimen Use of specimen threshold value Use of marker pen threshold and specimen threshold values Remove coverslip function ON OFF Use of scan everything within marked area function ON OFF Filling of holes ON OFF Use of removing specks according to the set v...

Page 64: ... effects damaging to health Decontaminate infectious materials after use Dust specks and humidity may decrease or harm the operating capability of the device Protect it with dust cover when not in use Switch the power off before covering it Avoid touching the lamp bulb Wait for about 15 minutes to allow the bulb to cool down and disconnect the power plug before changing the bulbs Defective devices...

Page 65: ... for the Pannoramic products If any problem arises that does not have solution in this manual contact the service personnel The product does not start Power supply is interrupted 1 Make sure that the 24V external power supply is on 2 Check cable connections Control software has stopped Name of currently selected slide for digitization already exists and is open in Pannoramic Viewer Close virtual s...

Page 66: ...eplacing the halogen lamp Virtual slide is improperly focused almost across its entire surface Focus range enabled Disable Limiting focus range Poor image quality Imaging error for example chromatic aberrations Contact 3DHISTECH Service for support Warning Do not modify the camera driver settings as the default factory settings give the most accurate images Virtual slides show poor stitching in Pa...

Page 67: ...tended focusing option if not required Z stack option is activated Deactivate the Z stack option if not required Image is out of focus Focus range limit is activated Deactivate the Limiting focus range option Focus range setting is incorrect Change focus range setting All individual images show a shading effect Algorithm failed to find empty image fields Capture a reference image with a blank slid...

Page 68: ...o replace the halogen lamp or to perform other maintenance work and to interrupt a running digitization procedure because of slide jamming the protective cover of the Pannoramic DESK II need to be removed in order to eliminate the problem Before removing the cover exit the control software switch off the 24 V power supply and disconnect the power cables Caution Only instructed persons or service t...

Page 69: ...rt of the cover must be pushed over the top part of the blue basic unit 7 5 To lock the cover in place insert the screws again and fasten them Caution Make sure that no cables are squeezed as you replace the cover s 5 3 2 Removing a slide If a slide got jammed or has suffered breakage during the process of digitization stop the process and if possible move the slide holder to a position where the ...

Page 70: ...Warning If the slide is broken make sure to remove all pieces completely from the device The edge of the broken area of the slide is sharp and it can lead to injury and contagion Broken slide can be removed only by a well trained person 6 Swing fixing lever 3 back to the right towards the holder 7 Reattach the cover see section 5 3 1 Removing the protective cover 8 Connect the power cable of the p...

Page 71: ...persons can carry the base unit by holding it by its base plate You can find the weight data of the base unit in section 6 Technical Data Caution Be sure that the scan camera is not moved during transportation To transport Pannoramic DESK II do the following 1 Turn power supply off and detach all cables 2 Transport Pannoramic DESK II 3 After finishing transportation reestablish cable connections C...

Page 72: ...upply approx 550 g Ambient conditions Transportation shipment packed Permissible ambient temperature 40 C to 70 C Storage Permissible ambient temperature 0 C to 60 C Permissible relative air humidity non condensing 75 Operation Permissible ambient temperature 10 C to 35 C Permissible relative air humidity 75 Altitude at installation site max 2000 m Atmospheric pressure 500 hPa to 1060 hPa Degree o...

Page 73: ... Power consumption of the complete device 75 VA 24V external mains power supply unit Intended site closed room facilities Electrical protection class I Internal protection degree IP 20 Electrical safety in compliance with UL 2601 1 TÜV EN 60601 CE EN60601 1 2 2001 Over voltage category II Radio interference suppression in accordance with CISPR 11 EN 55011 class B Line input voltage 100 V AC to 240...

Page 74: ...gh SNR mode Pixel resolution with 20 objective and 1 6 C mount adapter 0 24µm Connection type CoaXPress 4 lanes Point Grey Grasshopper 3 GS3 U3 51S5M M Global Shutter Type CMOS Sony IMX250 Pixel size 3 45µm 3 45µm Camera resolution 2448 H 2048 W pixel Frames rates 75 fps Bit depth 10 bit 12 bit Pixel resolution with 20x objective and 0 63x C mount adapter 0 27 µm Pixel resolution with 20x objectiv...

Page 75: ...erating System Microsoft Windows 7 SP1 64 bit EN CPU Intel Core i5 4590 RAM 8 GB DDR3 1600 Hard disk 1 TB Optical drive DVD RW Display 96 dpi Graphics SXGA 1280 1024 True color Input devices keyboard mouse Ports 2 x Fire Wire 1394 A 1394 B 4 x USB 2 0 GB Ethernet June 24 2016 Rev 1 3DHISTECH Ltd 75 76 ...

Page 76: ...Focus 12 18 43 46 49 53 54 55 63 66 67 H Halogen lamp 66 68 69 75 M Marker 45 48 63 P Pannoramic 6 Pannoramic DESK II 6 8 9 10 12 13 15 17 19 23 25 26 65 68 72 73 Pannoramic Viewer 18 54 65 66 Preview 20 22 24 26 31 40 44 45 46 47 48 49 51 52 54 76 R RGB 57 S Scan 12 14 17 31 40 41 42 43 44 45 46 47 48 54 55 56 58 63 66 72 76 Scan To CaseCenter 44 Scan To Server 60 61 62 Scanner 6 31 46 ScanToServ...

Reviews: