background image

Condenser:

 Abbe condenser N.A.1.25 with iris 

diaphragm, filter holder and blue filter, focussed 
via rack and pinion drive 

Dimensions: 

328 x 214 x 394 mm³ approx. 

Weight: 

6.1 kg approx.

 

 
 

3. Unpacking and assembly 

The microscope is packed in a molded styro-
foam container.  

 

Take the container out of the carton remove 
the tape and carefully lift the top half off the 
container. Be careful not to let the optical 
items (objectives and eyepieces) drop down.  

 

To avoid condensation on the optical com-
ponents, leave the microscope in the original 
packing to allow it to adjust to room tem-
perature.  

 

Using both hands (one around the pillar and 
one around the base), lift the microscope 
from the container and put it on a stable 
desk.  

 

The objectives will be found within individual 
protective vials. Install the objectives into the 
microscope nosepiece from the lowest 
magnification to the highest, in a clockwise 
direction from the rear.  

 

Put the head onto the top of the stand and 
tighten the head-lock-screw. Insert the eye-
pieces into the tube. 

 
 

4. Operation 

4.1 General information 

 

Set the microscope on a level table. 

 

Place the object to be observed in the centre 
of the specimen stage and clamp it to the 
object guide. 

 

Connect the mains cable to the net and turn 
on the switch to get the object illuminated. 

 

Make certain that the specimen is centered 
over the opening in the stage.  

 

To obtain a high contrast, adjust the back-
ground illumination by means of the iris dia-
phragm and the variable illumination control. 

 

Adjust the interpupillary distance so that one 
circle of light can be seen. 

 

Make the necessary eyepiece dioptre ad-
justments to suit your eyes. 

 

Rotate the nosepiece until the objective with 
the lowest magnification is pointed at the 
specimen. There is a definite “click” when 
each objective is lined up properly. 

NOTE: 

It is best to begin with the lowest power 

objective. This is important to reveal general 

structural details with the largest field of view 
first. Than you may increase the magnification 
as needed to reveal small details. 

When 100x 

(oil) objective is chosen, objective oil must be 
dripped onto the slide. 

To determine the magnification at which you are 
viewing a specimen, multiply the power of the 
eyepiece by the power of the objective. 

 

Adjust the holding brake to give a suitable 
degree of tightness in the focusing mecha-
nism. 

 

Adjust the coarse-focusing-knob which 
moves the stage up until the specimen is fo-
cused. Be careful that the objective does not 
make contact with the slide at any time. This 
may cause damage to the objective and/or 
crack your slide. 

 

Adjust the fine-focusing-knob to get the im-
age more sharp and more clear. 

 

Colour filters may be inserted into the filter 
holder for definition of specimen parts. 
Swing the filter holder out and insert colour 
filters. 

 

Use the knobs of the mechanical stage to 
move the slide side-, back- and forwards. 
The vernier provides accurate location of the 
specimen area. 

 

Always turn off the light immediately after 
use. 

 

Be careful not to spill any liquids on the mi-
croscope. 

 

Do not mishandle or impose unnecessary 
force on the microscope. 

 

Do not wipe the optics with your hands. 

 

Do not attempt to service the microscope 
yourself. 

 

4.2  Changing the lamp and fuse 

4.2.1 Changing the lamp 

 

Turn off the power switch, unplug the mains 
plug and let the lamp cool down to avoid be-
ing burnt. 

 

For safety reasons, remove the eyepiece. 

 

To change the lamp lay the microscope on 
its back to reach the lid on the bottom side. 

 

Loosen screw C of the lamp socket and 
push it outwards so that it is in the position 
shown in Fig.1. 

 

Loosen screw A and open the cover. 

 

To remove the halogen lamp, use a cloth or 
similar material. Do not touch the bulb with 
the bare hand. 

 

Lift out the halogen lamp and replace it with 
a new one. 

 

Close the cover and secure it with the 
screw. 

Summary of Contents for 400 1003275

Page 1: ...allstativ Stativarm fest mit Fuß verbunden Fokussierung über beidseitig am Stativ angebrachte koaxiale Stell knöpfe für Fein und Grobtrieb mit Kugellager und Feststellbremse einstellbarer Anschlag zum Schutz der Objektträger und Objektive Fokussierbereich 15 mm Einteilung der Feinfo kussierung 0 002 mm Tubus Binokularer Siedentopf Kopf Schräg einblick 30 Kopf um 360 drehbar Augenab stand zwischen ...

Page 2: ...stärkeren Vergrößerung zur Betrachtung feinerer Details erfolgt durch Drehen des Revol vers bis zum gewünschten Objektiv Bei Ver wendung des Objektivs 100x muss Öl auf den Objektträger gegeben werden Die Stärke der Vergrößerung ergibt sich aus dem Produkt des Vergrößerungsfaktors des Okulars und des Objektivs Mit der Feststellbremse geeignete Span nung des Fokusiersystems einstellen Mit dem Triebk...

Page 3: ... mit der Staubschutzhülle abdecken Das Mikroskop keinen Temperaturen unter 0 C und über 40 C sowie keiner relativen Luftfeuchtigkeit über 85 aussetzen Vor Pflege und Wartungsarbeiten ist immer der Netzstecker zu ziehen Zur Reinigung des Mikroskops keine ag gressiven Reiniger oder Lösungsmittel ver wenden Objektive und Okulare zum Reinigen nicht auseinander nehmen Bei starker Verschmutzung das Mikr...

Page 4: ......

Page 5: ...on Stand Robust all metal stand with arm perma nently connected to the base Focussing by means of separate knobs for coarse and fine adjustment located on either side of the stand and operated by rack and pinion drive with ball bearings and retaining lever adjustable stopper for protecting the object slides and objective Focus range 15mm resolution of fine focussing adjustment 0 002 mm Tube Binocu...

Page 6: ...operly NOTE It is best to begin with the lowest power objective This is important to reveal general structural details with the largest field of view first Than you may increase the magnification as needed to reveal small details When 100x oil objective is chosen objective oil must be dripped onto the slide To determine the magnification at which you are viewing a specimen multiply the power of th...

Page 7: ...d disposal Keep the microscope in a clean dry and dust free place When not in use always cover the micro scope with the dust cover Do not expose it to temperatures below 0 C and above 40 C and a max relative humid ity of over 85 Always unplug the mains plug before clean ing or maintenance Do not clean the unit with volatile solvents or abrasive cleaners Do not disassemble objective or eyepieces to...

Page 8: ......

Page 9: ...ort Bâti support entièrement métallique Grande robustesse Mise au point rapide par pignon crémaillère à roulement à billes fine par vis micrométrique à commandes coaxiales de précision avec frein de serrage Butée de pro tection réglable Mise au point 15 mm gradua tion de la mise au point fine 0 002 mm Tube Tête de type Siedentopf binoculaire inclinée à 30 tête orientable à 360 écarte ment interpup...

Page 10: ...f 100x mettre de l huile sur le support de l objet La force d agrandissement résulte du produit du facteur d agrandissement de l oculaire et de l objectif Avec le frein réglez la tension appropriée du système de focalisation Le bouton de mise au point rapide permet d obtenir une image nette de l objet encore flou veiller à ce que l objectif ne touche pas le support d objet risque d endommagement E...

Page 11: ...érieures à 40 C ainsi qu à une humidité relative de l aide supérieure à 85 Avant d effectuer des travaux d entretien et de maintenance retirer toujours la fiche secteur Pour le nettoyage du microscope ne pas utiliser de nettoyants ni de solvants agressifs Pour le nettoyage ne pas démonter les objectifs ni les oculaires S il est fortement encrassé nettoyer le mi croscope avec un chiffon doux et un ...

Page 12: ......

Page 13: ...00 volte Stativo Robusto stativo completamente in me tallo braccio dello stativo fissato saldamente alla base messa a fuoco attraverso manopole coassiali per regolazione macrometrica e mi crometrica separate poste ai lati dello stativo con cuscinetto a sfere e freno di arresto battuta regolabile per la protezione del portaoggetti e degli obiettivi Campo di messa fuoco 15 mm suddivisione della mess...

Page 14: ...ettivo con l ingrandimento mi nimo fino a portarlo sul percorso dei raggi luminosi Il raggiungimento della corretta posizione viene segnalato dallo scatto dell obiettivo Nota È opportuno cominciare con l ingrandimento minimo per poter riconoscere dapprima i dettagli macroscopici delle strutture Il passaggio a fattori di ingrandimento maggiori avviene attraverso la rotazione del revolver fino all i...

Page 15: ...tilazione C viti di fissaggio del portalampada D Portafusibili 5 Conservazione pulizia smaltimento Conservare il microscopio in un luogo pulito asciutto e privo di polvere Durante il periodo di non utilizzo coprire se pre il microscopio con la custodia antipolvere Non esporre il microscopio a temperature inferiori a 0 C e superiori a 40 né ad un umidità relativa superiore all 85 Prima di effettuar...

Page 16: ......

Page 17: ...tal ampliados entre 40 y 1500 veces Soporte Soporte de metal macizo brazo soporte de montaje fijo en el pedestal focalización por medio de ambos botones coaxiales para accionamiento burdo y fino con cojinete de bolas y freno de enclavamiento tope ajustable para protección del portaobjetos y de los objetivos Alcance de focalización 15 mm divisiones de la focalización 0 002 mm Tubo Cabezal binocular...

Page 18: ...ción regulable Gire el objetivo con el menor factor de ampliación hacia el haz de luz Un sonido de clic indica la posición correcta Nota Es mejor empezar con la ampliación más pequeña para en primer lugar reconocer los detalles más burdos de la estructura El paso a una ampliación mayor para la observación de detalles más finos se efectúa girando el portaobjetivos hasta llegar al objetivo deseado S...

Page 19: ...ilación C Tornillos se fijación del casquillo de la lámpara D Portafusible 5 Mantenimiento limpieza desecho El microscopio debe permanecer en un lugar limpio seco y libre de polvo Si se lo mantiene fuera de uso se debe cubrir siempre con la envoltura protectora contra el polvo No someta el microscopio a temperaturas bajo 0 C o sobre 40 C ni a una humedad relativa del aire superior al 85 Antes de r...

Page 20: ......

Page 21: ... microscópio binocular possibilita a observação em duas dimensões de objetos finas lâminas cortadas de plantas ou animais com um aumento de 40 até 1500 vezes Tripé Pé de apoio inteiramente metálico e robusto braço de apoio fixado no pé ajuste focal por meio de botões de ajuste fino e grosseiro coaxiais a ambos lados do pé de apoio movidos por rolamentos e com freio de posicionamento limite de apro...

Page 22: ...ve se efetuar o ajuste por meio da íris e da iluminação ajustável Girar a objetiva com o menor grau de aumento até a linha do raio luminoso O som de um clique indicará que a posição está correta Observação é melhor começar a observação com a objetiva de menor grau de ampliação de forma a reconhecer primeiro os detalhes estruturais maiores A passagem para um grau maior de aumento obtêm se girando o...

Page 23: ... eliminação Armazenar o microscópio num lugar limpo seco e sem poeira Quando não utilizado guarde sempre o microscópio na sua capa de proteção contra a poeira Não exponha o microscópio a temperaturas inferiores a 0 C ou superiores a 40 C assim como a uma humidade ambiente superior a 85 Antes de efetuar qualquer atividade de manutenção deve se sempre tirar o fio da tomada elétrica Não utilize produ...

Page 24: ......

Reviews: