1. Safety instructions   

The  tube  power  supply  unit  conforms  to  safety 
regulations  for  electrical  measuring,  control, 
monitoring  and  laboratory  equipment,  as  speci-
fied  under  DIN EN 61010,  section 1,  and  is  de-
signed  to  be  classified  as  protection  class I 
equipment.  It  is  intended  for  operation  in  a  dry 
environment suitable for the operation of electri-
cal equipment and systems.  

Safe  operation  of  the  equipment  is  guaranteed, 
provided  it  is  used  correctly.  However,  there  is 
no  guarantee  of  safety  if  the  equipment  is  used 
in  an  improper  or  careless  manner.  If  it  is 
deemed  that  the  equipment  can  no  longer  be 
operated  without  risk  (e.g.  visible  damage  has 
occurred), the equipment should be switched off 
immediately  and  secured  against  any  unintend-
ed use.      

In schools and other educational institutions, the 
operation  of  the  AC/DC  power  supply  unit  must 
be supervised by qualified personnel. 

 

Before putting the equipment into operation, 
confirm  it  is  compatible  with  the  local  mains 
voltage. 

 

Before  setting  starting  any  experiments, 
check  the  basic  unit  for  any  damage. In the 
event  of  any  malfunction/defect  or  visible 
damage,  switch  off  the  equipment  immedi-
ately and secure it against any unintentional 
use.      

 

The  instrument  may  only  be  connected  to 
the  mains  via  a  socket  that  has  an  earth 
connection.  

 

Never  cover  the  air  vents  at  the  rear  of  the 
housing. They are necessary in order to en-
sure  sufficient  circulation  of  air  required  for 
cooling  the  internal  components  of  the 
equipment.   

 

The 

equipment 

may 

only 

be 

opened/repaired  by  qualified  and  trained 
personnel.   

Warning:  The  equipment  supplies  DC  voltages 
in  excess  of  60 V  that  are  hazardous  to  come 
into contact with.   

 

Always,  switch  off  the  equipment  before 
making any modifications to the circuit. Only 
switch  on  the  equipment  when  the  experi-
mental set-up is complete.   

 

Use only safety experiment leads for making 
connections.   

 

Never touch or make direct contact with any 
connected or exposed leads, or the plates of 
a connected plate capacitor.  

 

2. Description  

The tube power supply unit is primarily designed 
for  operating  cathode  ray  tubes  (e.g.  a  narrow-
beam  tube  or  oscilloscope  demonstration  set). 
In  addition,  the  equipment  can  be  used  very 
effectively  for  conducting  experiments  on  the 
law  of  induction  and  on  charging  and  discharg-
ing of capacitors.      

The  power  supply  unit  is  capable  of  supplying 
voltage in the following ranges:  

1.  Stabilised  DC  (0-300 V)  for  acceleration  volt-
age in cathode ray tubes   

2.  Stabilised  DC  voltage  (0-50 V)  for  grid 
(Wehnelt) voltage in cathode ray tubes   

3. Stabilised DC voltage (4-12 V) for heater volt-
age in cathode ray tubes   

4.  Stabilised,  linearly  rising  or  falling  voltage  for 
quantitative analysis of induction processes with 
coreless  coils  and  for  charging  and  discharging 
of capacitors    

All  voltage  sources  are  floating  and  proof 
against  short  circuits  and  all  are  equipped  with 
an electronic voltage stabilisation system as well 
as  a  current  limiter.  In  the  event  of  a  thermal 
overload, the tube power supply unit is automat-
ically disconnected from the mains power supply 
and  is  automatically  reactivated  when  the  per-
missible temperature range has been attained.      
The  two  sockets  marked  “Timer”  are  equipped 
with  an  internal  relay,  which  is  closed  when  a 
rising or falling ramp is activated and is used for 
switching  on/switching  off  any  additionally  con-
nected  equipment  (timer,  oscilloscope,  x-t  plot-
ter, etc.) 

The tube power supply unit 1001011 is for oper-
ation with a mains voltage of 115 V (±10%), and 
the  1001012  unit  is  for  operation  with  a  mains 
voltage of 230 V (±10%). 

 

Summary of Contents for 1001011

Page 1: ... 4 Operation indicator for current limit 5 Output sockets 0 300 V constant linear rise fall 6 Common ground 7 Voltage regulator 0 50 V 8 Output sockets 0 50 V 9 Voltage regulator 4 12 V 10 Output sockets 4 12 V 11 PE socket 12 Sockets for timer 13 Regulator for rate of voltage drop 14 Rotary switch for increasing maintaining or reducing output voltage 15 Regulator for rate of voltage rise 16 ON OF...

Page 2: ... switch off the equipment before making any modifications to the circuit Only switch on the equipment when the experi mental set up is complete Use only safety experiment leads for making connections Never touch or make direct contact with any connected or exposed leads or the plates of a connected plate capacitor 2 Description The tube power supply unit is primarily designed for operating cathode...

Page 3: ...ge ramp Set the desired ramp using the rotary switch If the rotary switch is turned to the left from the STOP position the voltage at the output connectors of the DC 0 300 V voltage field increases linearly from 0 to the set final value of the ramp voltage If the rotary switch is turned to the right from the STOP position the voltage decreases from the set final value to 0 It is possible to halt a...

Page 4: ...ww 3bscientific com Technical amendments are possible Copyright 2015 3B Scientific GmbH 6 Sample experiments ı O On Off A PE PE 0 300 V 6 12 V 0 50 V H C W Fig 1 Experimental set up for filament tube ı O On Off Fig 2 Experimental set up for plate capacitor ...

Reviews: