© 2mag AG
Page 27 of 44
Version: 5.1
4.4.5 Control with HyperTerminal
The commands which are shown in spreadsheet 2 can be sent from the PC
to the controller with the Windows accessory program HyperTerminal.
Following HyperTerminal preferences allow to control at least 3 control
units and the summery of more than one command to a batch file.
bits per second: 9600 Baud
8 data bits, 1 stop bit, no parity bit
flow control: „XON / XOFF“ or „None“
Note: Flow control via hardware handshake is not working!
Minimum time distance between two commands: 40ms for one control
unit, minimum time distance 80ms for 3 control units.
HyperTerminal allows the adjustment of the time between two commands
via the line delay in the ASCII configuration.
For manual sending of commands via HyperTerminal the ASCII
configuration must be set which is shown in screenshot 4.