Caring for System
54
Cleaning your meter
To clean your meter, wipe the outside
with a soft cloth dampened with water
and mild detergent. Do not use alcohol
or another solvent to clean your meter.
Do not get any liquids, dirt, dust,
blood, or control solution inside the
meter through the test port.
Cleaning your lancing device and
clear AST cap
Clean and wipe the outside of the
lancing device with a soft cloth,
dampened with mild soap and water.
Wash the adjustable cap and clear AST
cap with mild soap. Do not immerse
the lancing device in any liquid.
Clinical accuracy
The accuracy of the 2in1. MICRO
Blood Glucose Monitoring System was
assessed by comparing blood glucose
results obtained by patients with those
obtained using a YSI 2300 Glucose
Analyzer.
System accuracy result for glucose concentrations
< 75 mg/dL (4.2 mmol/L)
System accuracy result for glucose concentrations
≥
75 mg/dL(4.2 mmol/L)
within ±5 mg/dL
(0.28 mmol/L)
80.0% (24/30)
within ±10 mg/dL
(0.56 mmol/L)
96.7% (29/30)
within ±15 mg/dL
(0.83 mmol/L)
100.0% (30/30)
within ±5%
63.5% (108/170)
within ±10%
94.7% (161/170)
within ±15%
99.4% (169/170)
within ±20%
99.4% (169/170)