WARNINGS AND PRECAUTIONS
IMPORTANT:
Additional
warnings and precautions
may appear in this user
manual.
WARNINGS
• The implant is not com-
pliant with the following
medical procedures. The
implant must be explanted
before receiving any of the
following treatments:
- MRI scan. The implant is
MR unsafe.
- Therapeutic ionizing
radiation induced close to
the implant (e.g. radiation
therapy for cancer).
- Therapeutic ultrasound
induced close to the
implant.
- Electrical current induced
close to the implant (e.g.
electro knife, electrocon-
vulsive therapy).
• The following medical pro-
cedures are safe to use with
the implant:
- Diagnostic ionizing radia-
tion (e.g. x-ray, CT).
- Diagnostic ultrasound.