ZyAura ZG1683R Скачать руководство пользователя страница 2

Ref.No.

072018

 

Specifications 

Method 

-Dual Beam NDIR (Non-Dispersive-Infrared)

 

Display

 - LCD Independent CO

2

, RH and Temperature readings   

Sample Method - 

Diffusion or flow through (50~200 ml/min). 

CO

2

 & Temperature Specification:

 

CO

2

 Specification:

 

Measurement Range 

0-3,000ppm 

Display Resolution 

1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm 

Accuracy 

Under 25°C: ±80ppm or ±5% of reading (

2000ppm); ±7% reading (>2000ppm)

 

Repeatability 

±20 ppm @ 400ppm 

Temperature Dependence 

Typ. ±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C 

Response Time 

About 2min for 63% of step change 

Warm-Up Time 

2 min. at 22°C 

Zone LED Display 

Green: <800ppm; Yellow: 800~1200ppm; Red: >1200ppm 

Temperature Specification: 

Temperature Range 

0°C~50°C (32°F~122°F) 

Display Resolution

 

0.1°C (0.1°F)

 

Display Options 

°C/°F 

Accuracy   

±1°C( ±2°F) When the fan blows to the device directly, the accuracy of temperature is ±1.5°C 

Response Time 

20-30 minutes (case must equalize with environment)

 

RH Specification:

 

Measurement Range 

20%~90% RH 

Display Resolution 

1% RH 

Accuracy 

±5% RH @ 23°C 

Response Time 

<5 min for 63% of step change 

Operating Conditions: 

Operating Temperature 

0°C~50°C (32°F~122°F)

 

Humidity Range 

0 ~95% RH non-condensing

 

Storage Conditions: 

Storage Temperature 

-20°C~60°C (-4°F~140°F) 

Power Supply:   

Power Supply 

USB or 5 VDC from external AC/DC adapter which is included in package (Use specified AC adapter only) 

DC Output Voltage   

5VDC 

AC Input Voltage & Frequency   

100~240VAC, 0.2A    50/60Hz 

 

Installation Instruction 

1. You can use ZG1683R on the table (please pull out the Housing Stand 

6

) or hang it on the wall.   

2. Please note the following when hanging it on the wall.   

Choose a suitable location to install ZG1683R, fix the screws to the wall.   
Please see

Dimension

section to check the distance between the screws.

 

Dimension 

 

                                                                                                                                                               

Fault Codes & Troubleshooting Guide 

This section includes a list of Frequently Asked Questions for problems you may encounter with the ZG1683R CO

2

 

Monitor. 

Fault Icon 

Description of the fault 

Suggested Actions 

 

The ambient temperature has exceeded the operating 
temperature range 0°C~50°C (32°F~122°F) 

This error will clear when the temperature returns to the range between 
0°C~50°C (32°F~122°F). 

 

 

EEPROM System Problem 

Please reconnect AC adapter to ZG1683R CO

2

 Monitor.   

If the “Err5, Err6” still appears, please contact the Service Department for 
further assistance. 

 

 