OPERATION
1.) Invisible light rays are continually emitted
from the faucet sensor.
2.) When the users hands come into range of
the sensors detection zone (8 approx.), the
beam is reflected back to the sensors reciever
and converted to a low voltage electrical signal
activating a solenoid valve.
3.) After the user removes their hands, the
absense of reflected light stops the low voltage
electrical signal, closing the valve. The circuit
automatically resets for the next user.
4.) If an object is in view continuously for thirty
seconds, the faucet will automatically shut off.
The faucet will remain off until the object is
removed. After the object is removed, the
faucet functions normally.
CARE AND CLEANING INSTRUCTIONS
DO NOT
use abrasive or chemical cleaners to clean faucets as
they will dull the luster and attack the chrome or special decora-
tive finishes. Use only mild soap and water, then wipe dry with a
clean cloth or towel. While cleaning the bathroom tile and floor,
the faucet and electronics should be protected from splattering
of water, cleaner, acids and cleaning fluids that can damage the
sensor faucet.
TROUBLESHOOTING GUIDE
PROBLEM
FAUCET WILL NOT
FLOW WATER
POSSIBLE CAUSE
Water supply not turned on.
Batteries not connected.
Cable connection is not
fully engaged.
Solenoid valve inoperative.
Sensor/electronics are
inoperative.
Filter is plugged.
Solenoid valve is dirty.
FAUCET CONTINUES
TO RUN FOR MORE
THAN 30 SECONDS
OR DRIPS
TO DIAGNOSE
Check supply stops.
Check connections.
Check cable connection.
Move hand in front of faucet to
see if solenoid valve is clicking.
Move hand in front of faucet
to see if solenoid valve is clicking.
See note in Step 4.
Shut off water supply. Remove
clean out cap.
Open the solenoid valve
and check for debris. Carefully
observe the way valve goes
together for reassembly.
REMEDY
Turn on water.
Reconnect batteries.
Remove battery, and disengage the
cable connection. Check cable
connection for broken pins and
reconnect cable.Check battery for
voltage level and reinstall.
If clicking, open solenoid valve and
clean it.
If not clicking, replace solenoid
valve.
Remove the battery and short
circuit the battery terminals with
allen wrench. This will reset the
electronics. Replace battery.
Clean the filter and reassemble.
Shut off water supply. Disassemble
solenoid and remove debris from
diaphragm and/or plunger tip.
Reassemble solenoid.
ZURN INDUSTRIES, INC.
¨
COMMERCIAL BRASS OPERATION
¨
5900 ELWIN BUCHANAN DRIVE
¨
SANFORD NC 27330
Phone: 1-800-997-3876
¨
Fax: 919-775-3541
¨
World Wide Web: www.zurn.com
In Canada: ZURN INDUSTRIES LIMITED
¨
3544 Nashua Drive
¨
Mississauga, Ontario L4V1L2
¨
Phone: 905-405-8272 Fax: 905-405-1292
AquaSense
is a registered trademark of Zurn Industries, Inc.
â
Printed in the U.S.A.
Form No. FV321, 9/02
SPECIFICATION
Voltage:
Sensor Range:
Operating Water
Pressure:
Ambient
Temperature:
Operational Water
Temperature:
Aerator:
6 VDC
(4 AA Alkaline)
7.5 Approximate
20-80 psi
32ºF to 122ºF (0ºC to 50ºC)
35°F to 140°F (2°C to 60°C)
2.0 GPM Vandal-Resistant (Standard)