MONITOR MODE (
Grey Zone
)
LEAD
- Change pressing lead. Circulates through Lead I-III and PADS
SIZE
- Change to make the waveform tracing bigger or smaller
ALARM SUSPEND
Default alarm volume is Off
•
Alarms On
– Press Alarm Suspend button to turn the alarms on
Temporary Silence -
Press again and it will be temporarily disabled the alarms (15 sec for lethal
and 90 sec for non-lethal rhythms)
Alarms Off -
Press and hold for 3
-
4 seconds and the alarms will be SILENCED until otherwise
activated by the user. (an x will appear next to alarm icons)
If the patient is in VFib, Vtac, or Asystole, the alarms will automatically sound after 5 seconds.
RECORDER
- Starts or stops the paper recorder at any time
SOFT KEYS
NIBP
(the button with the picture of an arm, bottom left)
Press once for one (1) NIBP reading. Press again to abort reading.
Press and hold for 2 Seconds for
STAT
mode (as many NIBP measurements as
possible in 5 minutes)
Options
QRS Vol On
- Turns tone on or off
Low Bright
- Adjust Screen Brightness
•
Traces
- Waveforms you can bring up. (Options——Traces——Trace 2,3)
•
More
- Reveals
“Clock Set”
to change time on display
Param
EtCO2 Setup
(
ONLY DONE ONCE
when a new sensor is connected to this R Series)
1. Insert EtCO2 Single Patient Airway Adapter
2. Press the “Param” Key
3. Press the “EtCO2” Key.
4. Press the “Zero” key. This will take about 10 seconds and will
display “Zeroed” when done.
NIBP Setup
1. Press the “Param” Key
2. Press the “Setting” Key
3. Press Auto Interval
4.
Set time for Auto Interval (default 5 minutes)
Code Marker
Varies with where the dial is positioned (Defib, Pacer, Monitor)
Allows you to time stamp different clinical actions and retrieve later on
Report Data
(Printing and Data Transfer)
o
Print Chart:
ECG strips for a specific time or all activated events
o
Print Range:
Select and print specific event
o
Print Log
: Chronological list of all interventions since the device has been on for the
current code and previous codes that are stored
•
TEST LOG
– Log of up to the last 1000 tests performed.
•
TRANSFER MODE
– Print to WiFi
Alarms
Set highs and lows for different parameters.
The alarms will be defaulted back once the device is off for 10 Seconds.