Propaq MD Error Messages
9650-0807-01 Rev. C
Propaq MD Service Manual
57
TEMP DEVICE FAILURE
Temp software detected a
problem with the
temperature data that is
most likely a hardware issue.
Replace the main board.
TEMP PROBE1 (or
PROBE2) SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.
TRANSMISSION FAILED
The communication
operation failed. This
message displays until a
new operation is started.
To correct the problem,
verify the following:
•
Wireless communication
is enabled on the
Propaq MD unit.
•
The Wi-Fi settings are
correct in the
Communications setup
menu.
•
The ZOLL server is
configured correctly.
•
Your cellular device is
configured correctly.
•
The Propaq MD unit is
within range of the
wireless server.
WAN CONNECTION
UNAVAILABLE
The Propaq MD unit was
unable to establish a WAN
connection.
To correct the problem,
verify the following:
•
Wireless communication
is enabled on the
Propaq MD unit.
•
The expected means of
coummunication is
enabled (for instance, if
using a cell modem,
make sure the USB cell
modem is set to
Enabled
).
•
(Bluetooth) It is paired
with the Propaq MD unit.
•
(Wi-Fi) The correct Wi-Fi
access point profile is
selected.
•
(Cellular modem) It is
properly connected to
the Propaq MD unit and
that the USB icon on the
display is illuminated.
Error Message
Explanation
Technical Action
Содержание Propaq MD
Страница 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Страница 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Страница 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Страница 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Страница 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Страница 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Страница 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Страница 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Страница 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Страница 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Страница 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Страница 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Страница 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Страница 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Страница 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Страница 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Страница 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Страница 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Страница 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Страница 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Страница 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Страница 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Страница 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Страница 142: ...This page intentionally left blank...