Preventive Maintenance Test
9650-0308-01 Rev. A
ZOLL AED Plus Service Manual
1–7
Turn off the AED Plus unit.
Set the analyzer to NSR (normal sinus rhythm).
2.5
Turn on the AED Plus unit.
!
UNIT OK
!
ADULT PADS
!
DON’T TOUCH PATIENT,
ANALYZING
.
!
NO TREATMENT ADVISED
(or
NO SHOCK ADVISED
)
These three prompts are optional:
—
OPEN AIRWAY
—
CHECK BREATHING
—
GIVE TWO BREATHS
!
CHECK CIRCULATION
(or
CHECK PULSE
)
!
IF NO CIRCULATION START CPR
(or
IF NO PULSE START CPR
)
!
!
2.6
Lift the puck that is attached to the
universal adapter cable. Move the
puck up and down very slightly
(no more than 1 inch or 2 cm) and
continue to do so.
If the unit has a display, prevent
the bar on the chest compression
gauge from extending into the
area between the lower two lines.
•
Unit emits spaced tones as a guide
for performing chest compressions
at the proper rate.
!
PUSH HARDER
!
!
2.7
Increase the vertical motion of the
puck (about 2 to 3 inches or 5 to
8 cm).
If the unit has a display, make
sure the bar on the chest
compression gauge extends into
the area between the lower two
lines.
•
Unit emits spaced tones as a guide
for performing chest compressions
at the proper rate.
!
GOOD COMPRESSIONS
The following prompt is optional:
—
IF NO CIRCULATION CONTINUE
CPR
(or
IF NO PULSE CONTINUE
CPR
)
!
STOP CPR
!
!
Test
Action
Expected Result(s)
Pass
Fail
Содержание Aed Plus
Страница 1: ...Service Manual 9650 0308 01 Rev A March 2005...
Страница 4: ...TABLE OF CONTENTS ii www zoll com 9650 0308 01 Rev A...
Страница 10: ...PREFACE viii www zoll com 9650 0308 01 Rev A...
Страница 20: ...CHAPTER 1 MAINTENANCE TESTS 1 10 www zoll com 9650 0308 01 Rev A...
Страница 26: ...CHAPTER 2 TROUBLESHOOTING 2 6 www zoll com 9650 0308 01 Rev A...
Страница 30: ...A 4 www zoll com 9650 0308 01 Rev A...