Physical Inspection
9650-0308-01 Rev. A
ZOLL AED Plus Service Manual
1–3
1.0 Physical Inspection
Use this procedure to ensure that the unit shows no signs of damage or excessive wear.
Tools Needed
None.
Test Setup
None.
Test Procedure
Observe the following:
Pass
Fail
1.1
Is the unit clean?
!
!
1.2
Are all signs of wear reasonable and not excessive?
!
!
1.3
Are the top panel and housing undamaged and free of cracks?
!
!
1.4
Are input connectors clean and undamaged?
!
!
1.5
If electrodes are preconnected, is the cable free from damage, are the pads
sealed within the packaging, and are the electrodes within the expiration
date shown on the packaging?
!
!