Automated External Defibrillator (AED) Guidelines 

 

 

A

th

le

ti

cs

 

 

Troubleshooting 

Problem 

Recommended Action 

Self-test failed 

Manually test by pressing and holding the ON/OFF button for 
more than 5 seconds. If unit fails test again, remove from service. 

“Change batteries” prompt 

Replace all batteries at the same time. 

Red “X” in Status Indicator 
window OR beeping noise when 
unit is OFF 

Perform manual test. Check to see if cable is attached properly to 
unit. Replace batteries. If unit still does not operate correctly, 
remove from service. 

Red “X” in Status Indicator 
window when unit is ON 

Power cycle the unit. If Red “X” is still present in Status Indicator 
window, remove unit from service. 

 
Scheduled Monitoring / Maintenance 

1.

 

Check the readiness indicator daily. 

2.

 

Check the AED according to the manufacturer’s recommended schedule and 
documented (Attachment B). 

3.

 

Check the Integrity of the case: 

a.

 

Examine the molded case of the AED for any visible signs of stress; and, 

b.

 

Contact manufacturer’s customer service if the case shows signs of stress. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Legal References: 

HB 1433 (Public Law #2009-424), HB 914 (Public Law #2012-198), SB 
1269 (Public Law #2000-113), 
SB 118, 29 CFR 1910.1030, P.L. 104-210 

Cross Reference: 

OSHA AED Standards; Zoll AED® Plus

 

product manual, 2015 AHA/ERC 

Guidelines AED Defibrillation CPR; NRMPS Bloodborne Pathogens Exposure 
Control Plan

 

Forms: 

NRMPS Accident Report Form, Medical Incident (9-1-1) Form, AED 
Maintenance Check Form, AED Training Log  