April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
22-1
Maintenance and Storage
22. Maintenance and Storage
The following table shows the recommended maintenance procedures for the
Nightingale PPM3 monitor and its accessories. The Nightingale PPM3 monitor
does not require periodic recalibration – with the exception of the Oridion
Microstream® CO
2
module (see the PPM3 Calibration Procedures after the
Functional Tests). However, it is a good idea to check that the monitor is in good
working order, as described in the following table. These functional tests should
be done every 12 months, and they can be performed by clinicians or qualified
service personnel. The following equipment is needed to perform these
procedures:
•
ECG/Respiration Simulator (e.g., Fluke Biomedical MPS-450)
•
SpO
2
Simulator (e.g., Fluke Biomedical Index-2)
•
Blood Pressure Simulator (e.g., Fluke Biomedical BP-Pump2)
•
Safety Analyzer (e.g., Fluke Biomedical 505 Series)
•
Oridion Microstream® calibration kit from Air Liquide (Scott
Medical); Part Number T4653ORFCD – including a canister of 5%
CO
2
, 21% O
2
, with the balance N
2
along with a T-piece connection
and a calibration Filterline.
22.1. PPM3 Functional Tests
PPM3 Function
Procedure
Mechanical Integrity
Check for cracks, abrasive edges and other signs of damage.
Knob
Verify menu navigation functionality by turning and pressing.
Front Panel Keys
PRINT key: verify that a Vital Signs Report is generated at the MPC’s printer.
ALARM SILENCE key: verify that alarm tone can be silenced.
NBP INTERVAL key: verify NBP Interval menu is displayed.
NBP START / STOP key: verify NBP measurement starts and stops.
STANDBY key: verify that unit enters Standby mode or displays Standby
Menu
SETUP key: verify that the top level Setup menu is displayed.
TRENDS key: verify that the Trends menu is displayed
MAIN SCREEN key: verify that the Main Screen is displayed
Power LED
Verify that the green power LED is illuminated on the front of the PPM3.
Battery Charging LED
Verify that the green charging LED is illuminated on the back of the PPM3
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......