ZKETECH
EBC-B20H User Manual
Copyright (C) 2006-2017 ZKETECH
4.3 Setting interface
Discharging test:
The first line shows step (1-5), mode and time (000-999 minutes, no limit when it’s set 0).
The second line shows discharging current and termination voltage.
Charging test:
The first line shows step (1-5), mode and termination current.
The second line doesn’t show.
Mode set includes DSC, CHG and END. If you set END the test stops and following steps
won’t work. Set CHG (step 1), DSC (step 2) and END (step 3) for one charging-discharging
circle – test won’t stop until the step 5 completes if you don’t set END.
The cursor sets automatically on the steps and you can move it to the option you want to
change. One press on the “SET” button to move one step to right. Rotate “SET” to left and
right to change the value. When the cursor is on the last value press the “SET” button to return
to the initial position. The value will change automatically if values set exceed limitation. Press
the “ON” button to switch to next setting option.
In setting mode, press “SET” knob for a long time to save data and return to testing interface.
When setting completes, press “ON” button to start test. A five minute interval is set
automatically between two steps to cool batteries and it shows WAT on the screen.
4.4 Setting examples
1) Set steps
Turn on the battery tester.
Connect battery and you can read the battery voltage on the testing interface on the
screen.
Press “SET” knob for more than two seconds to open the setting interface.
Set testing values.
Press “SET” knob for more than two seconds to return to testing interface.
Press “ON” button to start testing.
When testing completes rotate the knob to left and right to read test results.
Remove the battery and turn off the tester.