Page 2
INTRODUCTION
About
ZiXpress 32 Dx
system
Thank you for purchasing ZiXpress 32 Dx system
for your
laboratory. This guide contains important information
regarding the safe use of the ZiXpress 32 Dx system.
Please read this manual carefully, before starting to operate
the instrument for the first time, especially the Safety
Information.
If there is any question about how to install or operate it,
please contact our certified representatives / agents or email
our technical support center (
Technology Introduction
Principle
ZiXpress 32 Dx system utilizes the unique characteristic of
magnetic beads bonding with nucleic acid under a specific
environment to purify the target nucleic acid samples. The
purification protocol includes four basic steps: sample lysis,
nucleic acid binding, wash and elution. Using ZiXpress 32 Dx
system, the majority of nucleic acid purification processes will
be done automatically inside of ZiXpress 32 Dx system, it
dramatically reduce the human error and use the lab force
much more efficiently.
During the purification process, a 8-link rod comb, which is
attached to the tip comb rack, moves up and down repeatedly
to mix the reagents by creating turbulences in the 96 Deep-
Well Plate. When the permanent magnetic rods are insert into
the rod combs, the assembly lowers into the solution to
collect the magnetic beads at the bottom of the comb. The
whole assembly then move to a different Well and the
magnetic beads are then release by retracting the permanent
magnetic rods. ZiXpress 32 Dx system equips an exclusive
temperature control module that perfectly connect with the 96
Deep-Well Plate to enhance the lysis and elution efficiency.
Applications
ZiXpress 32 Dx system purifies DNA and RNA from a variety
of samples which are essential and broadly used in molecular
biology, genetic screening, sequencing, food safety, forensic,
etc.
Содержание ZiXpress 32 Dx system