Warnings
23
Ultrasound therapy
Do not conduct ultrasound therapy on patients with implants or any other
implanted electronic device unless the risk has been assessed and found to
be negligible.
Patients must not be connected to a radio-frequency surgical device at the
same time. This may result in burns.
Operation of the ultrasound device in the vicinity (e.g., within 1 m) of strong
electromagnetic fields (e.g., tomographs, X-ray or diathermy devices) may
interfere with the operation of the device. Please maintain a safe distance of
several metres.
Handle the ultrasound head carefully as rough treatment may alter its
properties. Do not bring the ultrasound head into contact with sharp or
pointed objects as the aluminium head is easily scratched.
Disinfect the ultrasound head with standard disinfectants after use.
The use of coupling agents other than the special ultrasound gel Sono plus
may damage the ultrasound head.
We recommend Sono plus by Zimmer MedizinSysteme.
Vacuum
Do not conduct treatment on patients with a tendency to haematoma unless
the risk has been assessed and found to be negligible.
This device is exclusively for use by qualified medical personnel. This
device may cause malfunctioning in or may interfere with the operation
of devices in its vicinity. It may be necessary to take action to avoid
interference such as using a different alignment, moving the device or
shielding it.