Manufacturer's EMC declaration
18
Page 38
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device en
Puls
version 2.0
is intended for use in the electromagnetic environment specified below. The
customer or user of the device en
Puls
version 2.0
should ensure that it is used in such environment.
Immunity Test
IEC 60601-Test
Level
Compliance Level
Electromagnetic Environment -
Guideline
Conducted Disturbances
induced by RF fields
according IEC 61000-4-6
3 V
0,15 MHz to 80
MHz
6 V in ISM Band
between 0,15 MHz
and 80 MHz
80% AM at 1 kHz
3 V
0,15 MHz to 80 MHz
6 V in ISM Band between 0,15
MHz and 80 MHz
80% AM at 1 kHz
In the vicinity of devices, bearing the
following symbol, interference is
possible:
Radiated RF EM fields
according IEC 61000-4-3
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz