Safety / Maintenance / Service
12.2 Service
12
Page 24
Air filter
An icon in the status bar indicates that the air filter needs to be replaced.
We recommend keeping a replacement air filter on hand in your medical
practice (see air filter information in this chapter).
Note
Replacing the air
filter
Note
Resetting the air
filter
Cryo 7 may not be operated without the Cryo 7 original air filter.
Only the original air filter from Zimmer MedizinSysteme GmbH may be used!
The air filter is located at the rear of Cryo 7.
Grasp with both hands from above through the handle.
Press the air filter downwards with your fingers at the indentations, pull it
forwards, and remove the air filter from the frame.
Insert the new air filter with the air filter frame first at the lower retaining
support. Press the air filter frame into the housing until it clicks into place. The
two arrows at the upper edge of the air filter must be pointing towards the
device. The retaining lugs on the air filter frame must click into place on the
side of the housing.
After the air filter is replaced, the operating time counter of the air filter must
be reset.
There are 2 options for performing this process:
Status bar
1.
Activating the "Air filter” icon in the status bar opens a window with
information on resetting the operating time counter.
Configuration menu, “user” service menu
2.
Activating the “Air filter operating hours reset”
button resets the
operating time counter.
Please dispose of the used air filter in household waste.