Safety and maintenance
15.1 Safety
15
Page 31
Cryo 6 is manufactured according to the DIN EN 60601-1 safety regulations.
As the manufacturer, Zimmer MedizinSysteme GmbH can only consider itself to be
responsible for the safety and reliability of the device if
x
the device is operated using a proper power outlet with earth contact and the
electrical installation complies with DIN VDE 0100 part 710,
x
the equipment is operated in accordance with the instructions for use,
x
extensions, readjustments or modifications are carried out only by persons
authorised by Zimmer MedizinSysteme,
x
the user has confirmed the functional safety and the proper operating condition
for him-/herself before use,
x
the device is operated only by properly trained personnel,
x
the device is not operated in areas at risk of explosion and / or a combustive
atmosphere, and
x
the device is immediately disconnected from the mains when penetrated by
liquid.
The device does not contain any parts that can be repaired by the operator.