22 |
Vanguard
®
ID Total Knee
Surgical Technique
Figure 32
A
B
Figure 33
Stem Preparation
Carefully impact the cruciate punch through the tibial
punch mask until fully seated (Figure 32).
Assemble the appropriate tibial tray trial cruciate stem
insert to the primary tibial tray trial (Figure 33 A and B).
Remove the tibial template and cruciate punch mask
and insert and impact the assembled primary tibial tray
trial into the tibia.
Trialing
The following bearing combinations are options based
on implant size (see the below sizing chart).
Note:
Medial and lateral bearings may differ in
thickness up to, but no greater than, 2 mm in order
to accommodate the soft tissue envelope of the
knee. Bearing articulation may also differ (see
bearing compatibility chart, page 24).
XP-CR
59
63
65
67
69
71
73
75
79
83
87
91
XP Bearings
59
63
71
79
87
AS Bearings
59
63
67
71
75
79
83
87
91
Locking Bar
59, 63, 65, 67
69, 71, 73, 75
79, 83, 87, 91
Sizing Chart