13 |
A.L.P.S.
Hand Fracture System
2.5 mm Plate Specifications
Plate Styles:
Straight
T
Y
Web
T/Y
Plate Width:
6.8 mm
27.6 mm
19.5 mm
17.4 mm
19.5 mm
Plate Thickness:
1.65 mm
1.65 mm
1.65 mm
1.65 mm
1.65 mm
Lengths:
90.2 mm
73.8 mm
71.8 mm
79.8 mm
55.5 mm
2.5 mm Locking Cortical Screw
• Large core diameter and shallow thread pitch
for bending and shear strength
• Self tapping tip minimizes the need for pre-tapping
and eases screw insertion
• Locking screw head designed to minimize screw
back-out and construct pullout
• When paired with the 2.5 mm Compression
Washer, it can be used as a compression screw
in the 2.5 mm plate compression holes
• Square drive
• Available in lengths of 8 – 28 mm
2.5 mm Non-Locking Cortical Screw
• Self tapping tip minimizes the need for pre-tapping
and eases screw insertion
• Square drive
• Available in lengths of 8 – 28 mm
2.5 mm Multi-Directional Locking
Cortical Screw (MDTP)
• CoCr Screws create new thread path in the plate
• Locking screw head designed to minimize screw
back-out and construct pullout
• Narrow shaft of MDTP is comparable to bending
stiffness of 2.5 mm locking screws
• Square drive
• Available in lengths of 10 – 28 mm
NOTE: With enough force it is possible to drive the 2.5 mm
MDTPs through the plate. Stop advancing the screw when
the head of screw is flush with the surface of the plate.
2.5 mm Compression Washer
• Converts only the 2.5 mm Locking Cortical Screw
into a compression screw
• Designed to be used in the compression holes
in the 2.5 mm plates
Содержание A.L.P.S.
Страница 1: ...A L P S Hand Fracture System Surgical Technique ...
Страница 2: ...2 A L P S Hand Fracture System ...
Страница 4: ...2 A L P S Hand Fracture System ...
Страница 6: ...4 A L P S Hand Fracture System ...
Страница 8: ...6 A L P S Hand Fracture System ...
Страница 17: ...15 A L P S Hand Fracture System T Y Web Y ...