6.8  Checking of parameters 

During operation, several parameters can be checked. 
 

 

 

Error search 

 

Wrong output signal (current/voltage) OUT (terminal 1 

– 2 – 3) 

Cause

 

The device is not configured correctly 

Remedy

 

Check commissioning 

 

22mA at output 0/4…20 mA, 11V at output 0/2…10 V (LED Power flashes quickly) 

Cause

 

Internal error of device  

Remedy

 

Reset by interrupting supply voltage. If the error cannot be patched by a Reset send 
back to factory for repair.  

 

Mode 

Normal state 

Содержание MU2000K

Страница 1: ...2 00 Page 1 8 www ziehl de Universal Measuring Transducer for Voltage and Current AC and DC Detailed operating manual see https www ziehl com en products view detail detail 110 1 Display and controls Set Button to navigate in menu Down Button to change Parameters LED Power Device active LED I Current input active LED U Voltage input active LED AC Alternate Current Voltage measuring active 2 Factor...

Страница 2: ...electrical units The equipment is built according to DIN EN and checked and leave the plant according to security in perfect condition If in any case the information in the instructions manual is not sufficient please contact our company or the responsible representative Instead of the industrial norms and regulations written in this instruction manual valid for Europe you must observe out of thei...

Страница 3: ... V 0 20 mA 80 V 120 V 0 A 5 A or 0 V 150 V 2 10 V 4 20 mA 0 V 300 V 0 V 600 V 6 2 Setting a predefined range Power off the device Press Button Set and keep pressed Power on the device and keep pressed the button Set After 5s flashes the green LED release button Set Parameterization input type LED ON flashes 1x Select with button Set the input type displayed by LED U DC voltage LED AC U AC voltage ...

Страница 4: ...ues will be discarded When scaling is started all parameters are reset to default In normal operating mode the selected input is displayed with LED 6 3 Diagram for setting a predefined range 6 4 Overview of scaling an arbitrary range Other ranges may be set by scaling of the input signal Range Output Input Zero point Full scale U AC DC 0 600 V 0 600 V 0 10 V 0 20 mA I AC DC 0 5 A 0 5 A or 2 10 V 4...

Страница 5: ...ton Down Parameterization input zero point LED ON flashes 2x Connect a signal at the input corresponding to the zero point Press button Down store of value green LED flashes 3 times quickly Parameterization input full scale LED ON flashes 3x Connect a signal at the input corresponding to the full scale Press button Down store of value green LED flashes 3 times quickly Parameterization output LED O...

Страница 6: ...s after app 5s when green LED starts blinking Parameterization System measurement time LED ON flashes 5x Select with button Set the System measurement time Number LEDs of flashes AC U I 1 x 20 ms 60Hz 17 ms 2 x 40 ms 60Hz 33 ms 3 x 80 ms 60Hz 67 ms 4 x 160 ms 60Hz 133 ms 5 x 320 ms 60Hz 267 ms 6 x 640 ms 60Hz 533 ms Press button Down End of parameterization LED ON lights permanently Signal input z...

Страница 7: ...ignal current voltage OUT terminal 1 2 3 Cause The device is not configured correctly Remedy Check commissioning 22mA at output 0 4 20 mA 11V at output 0 2 10 V LED Power flashes quickly Cause Internal error of device Remedy Reset by interrupting supply voltage If the error cannot be patched by a Reset send back to factory for repair Mode Normal state ...

Страница 8: ...tranded wire with insulated ferrules each 1 x 0 14 1 5 mm2 Torque 0 5 Nm 3 6 lb in Protection class housing IP 40 Protection class terminals IP 20 Mounting Snap mounting on 35 mm standard rail EN 60 715 or M4 screws Weight app 100 g Subject to technical changes 9 Housing Type K Dimensions in mm 1 lower part 2 upper part 3 bar 4 screw 5 holes for screw mount 19 8 19 8 35 4 22 5 75 0 82 5 110 0 1045...

Отзывы: