16 Maintenance
Attention :
Disconnect the instrument from the measuring circuit before
opening the instrument to replace the battery or the fuse !
16.1
Battery
Before initial start-up, or after storage of your instrument,
make sure that no leakage has occurred at the instrument
battery. Repeat this inspection at regular intervals.
If battery leakage has occurred, electrolyte from the battery
must be carefully and completely removed and a new battery
must be installed, before the instrument can be placed back
into operation.
If the " " symbol appears in the LCD display, you should
change the battery as soon as possible. You can continue to
take measurements, but reduced measuring accuracy may
result.
Replacing the
Battery(ies)
The housing base must be removed from the instrument in
order to replace the battery(ies).
Press the tab located beneath connector jacks with a test
probe, a banana plug or a similar object in the direction
indicated by the arrow as shown on the housing base, and
remove the base.
RA 10
Remove the battery from the battery compartment.
Insert two new 1.5 V mignon cells in accordance with the
polarity symbols in the battery compartment.
Place both battery cables between the cells before closing
the housing in order to prevent pinching of the cables.