Resistance and Continuity Measurements 

1. 

Insert the black test lead into the negative 

COM

 terminal and the 

 red test lead into the positive terminal. 

2. 

Set the function switch to the         

•))) 

 position.                                                              

3. 

Use the multifunction 

MODE

 button to select resistance. 

4. 

Touch the test probe tips across the circuit or component under  
test. It is best to disconnect one side of the device under test so  
the rest of the circuit will not interfere with the resistance reading. 

5. 

For Resistance tests, read the resistance on the LCD display. 

6. 

For Continuity tests, if the resistance is < 100

, a tone will sound. 

Diode Measurements

 

1.  Insert the black test lead banana plug into the negative 

COM

 jack and the red test lead 

banana plug into the positive diode jack. 

2.  Turn the rotary switch to the       

•)))

 position. 

3.  Press the 

MODE

 

button until “     “ appears in the display. 

4.  Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V. 

Reverse voltage will indicate “

OL

”. Shorted devices will indicate near 0mV and an open 

device will indicate “

OL

” in both polarities. 

 

 

 

Capacitance Measurements

 

WARNING:

 To avoid electric shock, disconnect power to the unit under test and 

discharge all capacitors before taking any capacitance measurements. Remove the 
batteries and unplug the line cords. 

1. Set the rotary function switch to the cap position. 
2. Insert the black test lead banana plug into the negative (COM) jack. 

Insert the red test lead banana plug into the positive (V)  jack. 

3. Touch the test leads to the capacitor to be tested. 
4. Read the capacitance value in the display 

Frequency or  % duty cycle measurements 

1. 

Set the function switch to the V position. 

2. 

Insert the black lead banana plug into the negative COM jack and the red test 
lead banana plug into the positive V jack. 

3. 

Select Hz or % duty with the 

Hz/%

 button. 

4. 

Touch the test probe tips to the circuit under test. 

5.

 

Read the frequency on the display.

 

 

Temperature Measurements 
WARNING: 

To avoid electric shock, disconnect both test probes from any source of 

voltage before making a temperature measurement. 

1. 

Set the function switch to TEMP. 

2. 

Insert the Temperature Probe into the negative (COM) and the V jacks, making sure 

to observe the correct polarity. 

3. 

Select 

C or 

F with the 

MODE

 button. 

4. 

Touch the Temperature Probe head to the part whose temperature you wish to 

measure. Keep the probe touching the part under test until the reading stabilizes (about 30 
seconds). 

5. 

Read the temperature in the display. The digital reading will indicate the proper 

decimal point and value. 

WARNING: 

To  avoid  electric  shock,  be  sure  the  thermocouple  has  been  removed  before 

changing to another measurement function 

 

Data Hold and Backlight 

To freeze the LCD meter reading, press the data hold button. The data hold button is 
located on the left side of the meter (top button). While data hold is active, the 

HOLD

 

display icon appears on the LCD. Press the data hold button again to return to normal 
operation. 

Note: The HOLD feature will activate when the Backlight is turned on. Press the HOLD key again 
to exit Hold. 

The backlight function illuminates the display and is used when the ambient light to too low 
to permit viewing of the displayed readings. Press the     (HOLD) button for one second to 
turn the backlight on and press the button a second time to turn the backlight off. 

 
Manual Ranging 

The meter turns on in the autoranging mode. Press the 

Range

 button to go to manual 

ranging. Each press of the range button will step to the next range as indicated by the units 
and decimal point location. Press and hold the 

Range

 button for two seconds to return to 

autoranging. Manual ranging does not function in the AC Current , Diode and Continuity 
check functions 

Battery Replacement 

1.  Remove the one rear Phillips head screw 
2.  Open the battery compartment 
3.  Replace the Requires one 

“9V” battery (NEDA1604, 6F22 006P)  

4. 

Re-assemble the meter

 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Red               Black     Black           Red 
Probe            Probe    Probe           Probe 

Forward test        Reverse test

 