TEMPERATURE MEASUREMENTS 

WARNING: 

To  avoid  electric  shock,  disconnect  both 

test probes  from any  source  of  voltage  before  making a 

temperature measurement. 

1.

 

If you wish to measure temperature in 

o

F, set the function 

switch  to  the 

o

F  range.  If  you  wish  to  measure 

temperature  in 

o

C,  set  the  function  switch  to  the 

o

range. 

2.

 

Insert  the  Temperature  Probe  into  the  negative  (-)  jack 

(COM) and the positive (+) jack (Temperature), making 

sure to observe the correct polarity. 

3.

 

Touch  the  Temperature  Probe  head  to  the  part  whose 

temperature  you  wish  to  measure.  Keep  the  probe 

touching  the  part  under  test  until  the  reading  stabilizes 

(about 30 seconds). 

4.

 

Read  the  temperature  in  the  display.  The  digital  reading 

will indicate the proper decimal point and value. 

WARNING: 

To  avoid  electric  shock,  be  sure  the 

thermocouple has been removed before changing to another 

measurement function. 

BATTERY TEST 

1.

 

Insert the black test lead banana plug into the negative 

COM

  jack  and  the  red  test  lead  banana  plug  into  the 

positive 

V

 jack. 

2.

 

Select  the 

9V

 

BAT

  position  using  the  function  select 

switch. 

3.

 

Connect the red test lead to the positive side of the 9V 

battery and the black test lead to the negative side of the 

9V battery. 

4.

 

Read the voltage in the display. 

 

Good 

Weak 

Bad 

9V battery:    >8.2V 

7.2 to 8.2V 

<7.2V 

 

CLAMP CURRENT TEST 

1.

 

Insert the clamp –on adaptor black banana plug into the 

negative 

COM

  jack  and  the  clamp  –on  adaptor  red 

banana plug into the positive

 uA, mA. A

 jack. 

2.

 

Select  the  clamp  current  DC  or  AC  position  using  the 

function select switch. 

3.

 

Press  the  trigger  to  open  jaw.  Fully  enclose  one 

conductor to be measured. 

4.

 

The meter LCD will display the reading. 

REPLACING THE BATTERY

 

WARNING

:  To  avoid  electric  shock,  disconnect  the  test 

leads  from  any  source  of  voltage  before  removing  the 

battery door. 

1.

 

When  the  batteries  become  exhausted  or  drop  below 

the  operating  voltage,  “BAT”  will  appear  in  the 

right-hand  side  of  the  LCD  display.  The  battery  should 

be replaced. 

2.

 

Follow  instructions  for  installing  battery.  See  the 

Battery Installation section of this manual. 

3.    Dispose of the old battery properly.

 

WARNING: 

To avoid electric shock, do not operate your 

meter  until  the  battery  door  is  in  place  and  fastened 

securely. 

BATTERY INSTALLATION 

WARNING

: To avoid electric shock, disconnect the 

test leads from any source of voltage before 

removing the battery door. 

1.

 

Disconnect the test leads from the meter. 

2.

 

Open  the  battery  door  by  loosening  the  screw  using  a 

Phillips head screwdriver. 

3.

 

Insert  the  battery  into  battery  holder,  observing  the 

correct polarity. 

4.

 

Put  the  battery  door  back  in  place.  Secure  with  the  two 

screws. 

WARNING: 

To  avoid  electric  shock, do not  operate  the 

meter  until  the  battery  door  is  in  place  and  fastened 

securely. 

NOTE

:  If  your  meter  does  not  work  properly,  check  the 

fuses  and  battery  to  make  sure  that  they  are  still  good  and 

that they are properly inserted.

 

REPLACING THE FUSES 

WARNING

:  To  avoid  electric  shock,  disconnect  the  test 

leads from any source of voltage before removing the fuse 

door. 

1.  Disconnect  the  test  leads  from  the  meter  and  any  item 

under test. 

2. Open the meter cover by loosening the screw on the cover 

using a Phillips head screwdriver. 

Содержание ZI-6940

Страница 1: ...When using the probes keep your fingers behind the finger guards on the probes Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock haza...

Страница 2: ...pushbutton SYMBOLS AND ANNUNCIATORS Continuity BAT Low Battery Diode DATA HOLD Data Hold AUTO AutoRanging AC Alternating Current or Voltage DC Direct Current or Voltage SPECIFICATIONS The instrument...

Страница 3: ...0k 2 0 of rdg 3 dgts Input Protection 250V dc or 250V ac rms Capacitance Auto ranging Range Resolution Accuracy 40 00nF 10pF 5 0 of rdg 7 dgts 400 0nF 0 1nF 4 000uF 1nF 3 0 of rdg 5 dgts 3 0 of rdg 5...

Страница 4: ...the indicator The indicator HOLD will be appear in the display 2 Press the DATA HOLD button to return to normal operation 3 Press the HOLD button for 2 seconds to BACKLIGHT button it will turn on or...

Страница 5: ...ion switch to the A position and insert the red test lead banana plug into the 10A jack 5 Press the MODE button until DC appears in the display 6 Remove power from the circuit under test then open up...

Страница 6: ...appears in the display 3 Insert the black test lead banana plug into the negative jack COM and the red test lead banana plug into the positive jack 4 Touch the test probe tips to the diode or semicond...

Страница 7: ...or AC position using the function select switch 3 Press the trigger to open jaw Fully enclose one conductor to be measured 4 The meter LCD will display the reading REPLACING THE BATTERY WARNING To avo...

Страница 8: ...er 4 Always use a fuse of the proper size and value 0 5A 250V fast blow for the 400mA range 20A 250V fast blow for the 10A range 5 Put the cover Insert the screw and tighten it securely WARNING To avo...

Отзывы: