CA103&CA102S Handy Multi-functions Calibrator Operating Instruction
30
3 Key in the temperature reading, the result will be shown as below:
Press ESC it will back to main screen.
8. Charging
Normal operation can be measured or output signal calibrator charge.
The battery have to be charged when it flat (Calibrator cannot turn on or
LCD flicker).
Please do the following:
1 Power adapter plugged into an AC outlet.
2 Connect the power adapter to the device
Green LED lights up while the Ni-MH battery pack is charged. Red
LED lights up while the Ni-MH battery pack is Fast charged.
indicate
Red LED
Green LED
Charging
time(Minute)
Charging
Light on
Flicker
0-5
Fast charging
Light on
Light on
About 360
trickle charging
Light on
Light off
About 360
3 When fast charging is completed (rapid charging indicator light
goes out) after then please keep the trickle charge 6 hours, to ensure that
the battery is fully charged.
4 When the device is fully charged, disconnect the power adapter and
the device, and then unplug the power adapter from the wall outlet.
Query
Cu50
Temp : 0000.00
℃
ITS-90 : 049.99
Ω
http://www.zhoyi.com