45
46
TROUBLESHOOTING
SPECIFICATIONS
Error
message
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
Warning
message
out shows
Out of measurement
range
E 03 shows
E 04 shows
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
The measurement
process does not
detect the pulse
signal.
Loosen the clothing on
the arm and then
measure again.
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.
Power supply
Battery powered mode:
6VDC 4*AAA batteries
AC adaptor powered mode: 5V 1A (Not included)
(Please only use the recommended AC
adaptor model).
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Pressure:
41°F-104°F (5°C-40°C) within±3mmHg(0.4kPa)
Pulse value:±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Digital LCD display V.A.2.36"X2.91"(60mm*74mm)
A temperature range of :+41°F to +104°F(+5°C to +40°C)
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water
vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa