Infinity
®
Orange
™
Operator’s Manual
14
2. DIRECTIONS FOR USE
The Battery
Running on Battery Power
Infinity Orange will run for 24 hours at a rate
of 125 ml/hr. A fully depleted battery takes ap
-
proximately 6 hours to fully charge.
When pump is running on battery power the
battery symbol will appear in the display (
Fig-
ure 2-17
).
The segments or blocks between E and F represent the fuel gauge of the battery.
Each block represents approximately 6 hours of charge. When each bar is half
spent, it will begin to flash, indicating approximately 3 hours of charge is left for
that bar. When the last block is gone, the battery symbol will blink to indicate there
is approximately 30 minutes of charge left. Display will flash LOW BATT every 3
seconds and pump will beep every 2 seconds to remind user of low battery condi
-
tion. When battery is fully depleted pump will turn off automatically.
NOTE: The battery life is an approximation based on a fully charged battery and a
flow rate of 125 ml/hr. Higher flow rates will run the battery down faster, while lower
flow rates will allow the battery to last longer. Battery life degrades over time.
Figure 2-17 Battery Power Indicator