![Zenith Therm-X AT Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/zenith/therm-x-at/therm-x-at_user-manual_929425016.webp)
15
IFU-001 Rev. 8
Thermal Garment Compression
For thermal garment compression, either constant or
intermittent pressure may be selected.
If constant pressure is selected, the thermal garment will
fill to the selected pressure, and remain at that pressure
for the duration of the treatment cycle.
If intermittent pressure is selected, the thermal garment will fill to the selected pressure, will hold and
then release the selected pressure in cyclic fashion for the duration of the treatment.
DVT Garment Compression
DVT treatment may be programmed with any thermal treatment cycle, or selected as an independent
treatment from the Select Treatment Cycle screen. If one DVT garment is being used, compression will
increase and release throughout the course of the treatment (one minute of pulsing compression, one
minute of rest). If two DVT garments are being used, treatment will alternate from one garment to
another and pressure will increase and release throughout the course of the treatment.
Pause Treatment
You may pause treatment at any time, using the Pause button on the bottom left corner of the Therm-
X touchscreen. Pausing treatment will release all garment pressure and stop warming or cooling the
thermal garment. You may resume treatment by pressing “Resume.”