13
5.4.7. L command
function
It can be set with or without latch circuit at the time of data input.
When the ZS-6222A is powered on, it is set to no latch circuit.
format
Lx
CR LF
: x is a number, "0" if no, "1" if yes.
return value
OK
CR LF
: Setting complete
NG
CR LF
: Setting error, wrong setting error
5.4.8. U command
function
This is used to set the signal setting for data output.
When the ZS-6222A is turned on, it is set to continuous output.
The pulse width at the time of pulse output is set by the P command.
format
Ux
CR LF
: x is a number, "0" for continuous output and "1" for pulse output.
return value
OK
CR LF
: Setting complete
NG
CR LF
: Setting error, wrong setting error
5.4.9. B Command
function
Positive and negative logic settings for the four ports.
When the ZS-6222A is powered on, all are set to positive logic.
This command should be performed when the ports are set to input.
format
Bx
CR LF
: x is a number, "0" for positive logic and "1" for negative logic.
return value
OK
CR LF
: Setting complete
NG
CR LF
: Setting error, wrong setting error