6
7
TABLE OF CONTENTS
SUBJECT SECTION
PAGE
8.
Maintenance
40
8.1
Battery Installation
40
8.2
Alkaline Battery Pack Installation
40
8.3
NiMH Battery Pack Installation
41
8.4
NiMH Battery Pack – General Information
41
8.5
Sensor Replacement
41
8.6
Removing and Adding a Sensor
42
8.7
Cleaning
42
8.8
Pump Inlet Filter Replacement
42
8.9
Sampling Hose Maintenance
42
8.10 Storage
43
8.11 Sensor and Battery Disposal
43
9.
Troubleshooting
44
9.1
Sensor Not Recognized (“XXX” Displayed)
44
9.2
“LEL FAIL” Message
44
9.3
Low Flow Alarm When Sampling Hose Attached
44
9.4
“Calibration Fail” Message
45
9.5
Instrument Will Not Turn On
45
9.6
Instrument is in Diffusion Mode but Pump
Comes On When Calibrating
45
Appendix A
Specifications
46
Appendix B
Turn-on Sequence
59
Appendix C
Mode Sequence, Standard Mode
61
Appendix D
Mode Sequence, User Set-up Mode
62
Appendix E
Alarm Modes
66
Appendix F
Charging Screens
68
Appendix G
Calibration Kits
69
Appendix H
Optional Accessories
70
Appendix I
Replacement Parts
70
Appendix J
Calibrating with Multiple Canisters
72
Appendix K
Calibration Gases
73
I.
PREPARATION FOR USE
1.1 UNPACKING
Check for these items:
•
Packaging (retain for possible future use)
•
Instrument with internal rechargeable battery pack
•
Calibration certificate
•
Standard accessories
:
–
AC battery charger
–
Alkaline battery holder
–
Dust filter
–
10-foot sampling hose with filter and water trap
–
Rubber boot with belt clip and neck strap
–
Instruction manual
1.2 CHARGING
CAUTION
Do not change or charge batteries in a hazardous location.
Do not try to turn on or otherwise operate this instrument until it is fully charged.
Use only the MPRO-1 AC charger, the MPRO-2 220 volt AC charger, or the MPRO-6
vehicular charger. Use of any other charger may cause permanent instrument damage and
will void any warranties. The charge period is dependent on the condition of the batteries,
but should not exceed about 4 hours for a completely depleted battery pack. Partially
depleted packs will be charged in a much shorter period of time.
To commence charging with the charger supplied with AC power, connect the charger plug
to the instrument port (See Figure 1) and instrument display should read “CHARGING
BATTERY”. When charging is completed, “CHARGE COMPLETE” is displayed.
The instrument is now fully charged and ready for use.
If charging is attempted without the rechargeable battery,
or with alkaline batteries, “CHECK BATTERY” is displayed.
M U L T I G A S D E T E C T I O N
Figure 1
Charging the Mi