background image

B 46-0008 e   12/06 

www.zeiss.de/stemi 

Adjusting the illumination  

 

 

 

 

Reflected-light illumination: 

 

Connect the light guide for reflected-light to the cold-light source 

(RL) and attach it to the stand plate or nosepiece (or carrier S). 
Alternatively clamp the ring illuminator to the single objective (not 
to the nosepiece). 

 

Switch on the cold-light source for reflected-light (RL) with rocker 

switch 

I

 

Align the light guide on to the specimen. 

 

Adjust the illumination intensity  on the cold-light source or use 

the two right hand buttons on S Y C O P . 

Transmitted-light illumination with transmitted-light unit S: 

 

Connect the light guide for transmitted-light (TL) to the cold-light 
source. 

 

For lower magnifications, insert the opal glass. 

 

To avoid dazzle from transmitted-light, swing the dazzle protector 
into position. 

 

Switch on the cold-light source for transmitted-light (TL) with 
rocker switch 

I

 

Adjust the illumination intensity  on the cold-light source or use 
the two left hand buttons on S Y C O P . 

 

The illumination levers on the transmitted-light unit allow the 
brightfield / darkfield to be adjusted as depicted alongside. To 
optimise the contrast with oblique light various intermediate 
positions can be employed (see explanation of symbols). 

Symbols for illumination control 

 

 

 

 

 

 

 

 Light beam exits vertically (reflector at 45° inclination) 

 

 

  

Inclining

 the light beam towards the observer (dazzle hazard!) 

 

Inclining

 the light beam away from the observer 

 

  Light beam exits centrally  

  

Translation

 of the light beam towards the observer  

 

  Diffuse 

light quality 

with the white reflector plate 

  Directed 

light quality 

with the mirror reflector plate 

 

 

 

Adjusting the stereomicroscope 

 

 

 

  The stereomicroscope should be connected and switched on. 

Focus and zoom can be controlled using H I P  or S Y C O P  . See section 

H I P  controls

 or 

S Y C O P   controls

 

Place a specimen on the insert plate (

3

). 

 

If the dioptric eye values are known, set the eyepieces (

2

accordingly. 

 

Select the viewing height by twisting the eyepiece support (

1

) by 

up to max. 120° and adjust it to the pitch of individual eyes, i. e. 
when looking in the eyepieces only an uninterrupted circle of light 
should be seen. 

 

Tilt the binocular part viewing angle and set it within the range  
5° - 45°. 

 

Open the aperture diaphragm using thumb wheel (

4

). 

 

When searching for objects, first set the zoom (

5

) to the lowest 

magnification. 

 

Focus on a small sharp detail in the centre of the image (

6

). 

 

Set the maximum zoom value (

5

) and adjust for exact focus on the 

detail (

6

). 

 

Then go back to the lowest zoom value (

5

) and correct any out-of-

focus separately for each eye, using the adjustable eyepieces (

2

(compensation for variations in the eye). 

 

Once the microscope has been compensated in this way, image focus 
will be retained over the entire zoom range.  

At the camera outlet of the Ergo phototube (

8

) a camera can be 

attached by a suitable adapter if desired. To change between visual 
observation and documentation, swing over the lever (

7

). The right 

hand beam path is switched over completely. 

 

 

 

 

Отзывы: