UNPACKING AND INSTALLATION 

SDA Stand for Stereomicroscopes 

 

Carl Zeiss 

 
 

 
 
B 46-0075 e   05/06 

UNPACKING AND INSTALLATION 

3.1 

Unpacking and installing the SDA 
stand 

The two standard basic components of the SDA 
stand are supplied packed in two packages. 

The SDA base plate is separately packed. 

The package containing the double boom 
additionally includes: 

− 

SDA column 

− 

Double boom with rotating/tilting head and 
column 32 

− 

Vertical adjustment retainer ring 

− 

Screws, lever and tools 

• 

For the installation of the SDA stand, remove all 
components from the packaging (verify that all 
components are present). 

 

CAUTION

 

When unpacking and installing the 
base plate, consider that it has a 
weight of approx. 30 kg. 

 

Two persons are needed to screw the column into 
the base plate: 

• 

One person should hold the base plate (up end) 
and the column and align these parts to each 
other. 

• 

The second person should screw the column 
(3-1/

1

) onto the base plate (3-1/

2

) with 

mounting screw (M16, 3-1/

3

) using an Allen key 

(SW 14 mm). The mounting screw should be 
fastened tightly. 

• 

Erect base plate with column and put it onto a 
stable support. 

 

Fig. 3-1 

Screwing the column onto the base 
plate 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание Stemi 508

Страница 1: ...SDA Stand for Stereomicroscopes Operating Manual...

Страница 2: ......

Страница 3: ...not covered by an update service Unless expressly authorized forwarding and duplication of this document as well as utilization and communication of its contents are not permitted Violations will ent...

Страница 4: ...e SDA stand 9 3 2 Installing Stemi mount and microscope body 11 4 Operation 12 4 1 Vertical adjustment of the double boom 12 4 2 Longitudinal adjustment of the double boom 13 4 3 Boom rotation about t...

Страница 5: ...warnings that must be observed by the user The following warning and information symbols are used in this manual CAUTION This symbol indicates a possible hazard to the user of the instrument CAUTION T...

Страница 6: ...teps be taken to minimize damage If notified of such a defect the manufacturer is obligated to rectify it at his discretion either by repairing the instrument or delivering an intact replacement No gu...

Страница 7: ...DESCRIPTION SDA Stand for Stereomicroscopes Carl Zeiss B 46 0075 e 05 06 7 2 DESCRIPTION 2 1 System overview...

Страница 8: ...ment range of double boom Horizontal in guiding direction max 330 mm Vertical on SDA column 425 mm Rotary on SDA column 360 Accessible object positions Swivel radius around SDA column measured up to A...

Страница 9: ...ols For the installation of the SDA stand remove all components from the packaging verify that all components are present CAUTION When unpacking and installing the base plate consider that it has a we...

Страница 10: ...r position After releasing the lever it will automatically snap in again by a spring mechanism CAUTION The double boom has a weight of approx 15 kg call on a second person for help if necessary Slip t...

Страница 11: ...With a short working distance of the supplementary lens system objective you may also slip the Stemi mount turned about over the column 32 to increase the spacing to the object and the base plate If...

Страница 12: ...1 1 may only be released in turn otherwise the double boom will slip down unsecured involving the risk of injury to the operator or damage to the microscope and the object to be examined 4 1 Vertical...

Страница 13: ...adjustment of the double boom CAUTION Double boom and weight of the microscope must be adjusted to each other to ensure stability of the setup and prevent the risk of injury to the operator or damage...

Страница 14: ...and clamp it tightly 4 1 3 Loosen the boom rotation clamp screw 4 1 1 and swivel the double boom to the desired position CAUTION Take care to prevent any collision with persons the object to be examin...

Страница 15: ...base plate Move the double boom up before if necessary Loosen clamp lever 4 1 5 on rotating tilting head 4 1 7 Incline column 32 with Stemi mount and microscope into the desired position Retighten the...

Страница 16: ...sible For further details please contact Specifications are subject to change B 46 0075 e 05 06 Carl Zeiss MicroImaging GmbH P O B 4041 37030 G ttingen Germany Phone 49 0 551 5060 660 Fax 49 0 551 506...

Отзывы: