CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Laser TIRF 3
Fuse Replacement
Carl Zeiss
02/2014
423683-7144-001
87
5.4
Fuse Replacement
Make sure the power cable is disconnected before changing any fuses.
Axio Observer, HBO, XBO, HXP, Tempmodul S1: see separate operating manual.
Laser Bench Rack and TIRF Controller: Do not contain fuses changeable from outside. Please contact the
Service Department of Carl Zeiss Microscopy GmbH.
5.4.1
Signal Distribution Box
The fuse compartment is combined with the
coupler plug and contains two fuses.
•
Switch off the instrument by means of the
power switch.
•
Disconnect the power plug.
•
Remove the fuse carrier (Fig. 5-1/
1
) by pulling it
towards you. You may use a small screwdriver,
if necessary.
•
Remove the fuses (Fig. 5-1/
2
) from the fuse
holder and insert new fuses (2x), type
T 1A/H/250 V
.
•
Push the fuse holder back into the fuse
compartment as far as it will go.
•
Connect the power cable.
5.4.2
Laser Bench Tower
•
Switch off the instrument by means of the
power switch.
•
Disconnect the power plug.
•
Pull the fuse holder (Fig 5-2/
1
) out of the fuse
compartment.
•
Replace relevant fuses (2x
T6.3 A/H/250 V
;
2
).
•
Insert fuse carrier.
•
Connect the power cable.
Fig. 5-1
Changing the fuses of the signal
distribution box
Fig 5-2
Changing the fuses of the Laser
Bench Tower