ZEISS SL Imaging Solution Quick Guide

Option 1: Eyepiece with crosshair (recommended)

This method applies to all types of slit lamps

1. 

Insert the eyepiece with crosshair into the right side of the  

binocular tube (this is where the camera is located).

2. 

Close the left eye and turn the diopter ring of the eyepiece  
until the crosshair appears sharp. Start from maximum "+"  
to avoid accommodation.

3. 

Ensure the crosshair and observed object are both in focus. 

4. 

Check the monitor image; it should now be in focus.

Good image

Crosshair and image in focus

Poor image

Crosshair in focus, but view out 
of focus » 

Focus slit lamp

Poor image

View in focus, but crosshair out 
of focus » 

Focus eyepieces

Issue: Observer can compensate a misalignment of the instrument, the camera cannot

How to align the slit lamp and camera correctly

Option 2: Monitor focus (no crosshair eyepiece  
or focus rod)

This method applies to all types of slit lamps

1. 

Attach an object (e.g., a business card) to the headrest.

2. 

Focus on that object using the highest magnification while 
looking only at the monitor.

3. 

Block the slit lamp by activating the quick action break.

4. 

Looking through the microscope, without moving the slit 
lamp, adjust your eyepieces (monocular).

Tips:

•  Keep in mind which beam path is recorded to avoid  

shadows and reflexes

•  Use the aperture to increase the depth of field
•  Use the optimal type of illumination
•  Big illumination angle for 3D information

•  For people who wear glasses, it is necessary to slide  

in eyecups

•  Reduce ambient light for better results
•  Accommodation is more pronounced among younger 

people (that’s why adjusting correctly, and Option 1,  
is important)