System enhancements
Requirements for integrating the device into an existing IT network
When connecting the device to an existing IT network, comply with the
following specifications and measures in accordance with IEC 60601,
Amendment 1:2012-08, section 14.13:
–
The device can be connected to an existing IT network in order to store
video and image data on a compatible network server.
–
The IT network to which the device is to be connected must have the
following features:
–
IPv4, static or dynamic address
–
The IP address range is arbitrary
–
The IT network to which the device is to be connected must be configured
as follows:
–
The firewall is open for inbound ports.
–
All data are transmitted unsecured from the device to the customer's
secured internal network.
–
Data is not intended to be transmitted from the device to external
networks.
–
Requirements in relation to network connection with the device:
–
The IT network has an Ethernet connection via RJ45, at least Cat.5,
100 Mbit/s.
–
The device has a network isolator in compliance with 60601-1,
and is therefore not connected to the mains power supply.
The operator is responsible for the topology of the IT network and for the flow
of information and the routing between the device, the IT network and other
devices.
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...