Requirements for connecting the network computer:
– You may need administrator rights to the network computer to
create a network share.
– You will need the Ethernet data of the archiving PC to configure
the network share.
– You can obtain the data required for the network share from your
IT Administrator.
• Enter the data and use "Apply" to activate the changes.
3
Streaming
You can stream live camera data to a network computer with little delay.
Depending on the receiver software, delays of less than 1 second are pos-
sible.
Prerequisites for Streaming:
– You need RTSP/RTP compatible receiver software in your compu-
ter of media device.
• Enter the data and use "Apply" to activate the changes.
4
Web control / Web interface*
The web interface with activated web control is a substitute remote con-
trol that uses a web browser.
When web control is activated, you can control the following functions
over the network at the computer:
– Manage patients and input or rename patients directly using the
– Delete files or patients
– Trigger the <Record>, <Capture>, <Live> and <Index> keys (see
– The image is displayed or the video is played via the DVI or YPbPr
monitor output, and not in the web browser.
If web control is deactivated, control is not possible. No header bar and
footer bar or notifications are displayed in the web browser (security func-
tion e.g. for universities).
Security recommendations for networking:
– You can deactivate the use of the network. Switch over to the UBS or dis-
connect the network cable.
– You can deactivate web control over the network.
Содержание S100/OPMI pico
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...