OPMI Lumera i
Safety measures
Controlling the surgical
microscope via CALLISTO eye
Any change in the focus and zoom setting via CALLISTO eye may lead to
unwanted movement at the surgical microscope during application (e.g.,
in the OP mode). Only personnel who have been specifically trained are
allowed to control the surgical microscope via CALLISTO eye (option).
Transport & Service
•
The system may be transported over long distances (e.g., during relocation,
return for repair, etc.) only in its original packaging or in special return
packaging. Please contact your dealer or ZEISS Service for this purpose.
•
If there is a failure and you cannot resolve the issue using information in
the chapter “What to do in the event of malfunctions”, attach an “Out of
order” sign to the system and contact ZEISS Service.
Safety inspection
•
To prevent any impairment of the system’s safety as a result of aging, wear,
etc., the organization operating the system must ensure, in accordance with
the applicable national regulations, that the regular technical safety checks
defined for this system are performed on schedule and to the stipulated
extent. The technical safety checks must only be performed by the
manufacturer ZEISS or persons authorized by ZEISS. The safety inspections
for the system should at least comprise the following points:
–
Availability of the Instructions for Use
–
Visual inspection of the device and accessories for damage, as well as
legibility of the labels
–
Leakage current test
–
Test of protective ground conductor
–
Function and wear test of the brakes
–
Function test of all switches, buttons, sockets, and indicator lamps of
the device
–
Function test of the manual mode without the patient, at least semi-
annually.
Medical electrical
systems
Any additional equipment you connect to medical electrical devices must
demonstrably comply with the applicable standards and directives (e.g.,
IEC 60950-1 for data processing equipment).
In addition, all configurations must meet the standard requirements for
medical electrical systems (see IEC 60601-1).
Anyone connecting additional equipment to medical electrical devices is a
system configurer and as such is responsible for compliance of the system
with the normative requirements for medical electrical systems.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...