What to do in the event of malfunctions
OPMI Lumera
®
i on floor stand
Version 9.2
Page 126
G-30-1720-en
Troubleshooting
For your safety
•
Many malfunctions are detected automatically. Therefore, pay attention
to the error messages displayed on the 5.7
″
control panel (see page 129).
•
If there is a failure you cannot repair using the instructions from the chap-
ter “What to do in the event of malfunctions”, attach a sign “Out of order”
to the system and inform ZEISS Service.
System malfunctions
Malfunction
Possible cause
Remedy
Reference
No function at all.
Power plug of stand not
inserted.
Plug in the power cord.
Power switch of stand not
switched on.
Press power switch.
Green indicator light in the
power switch must be lit.
Automatic circuit breaker
in the stand power switch
is being triggered.
Press the power switch
again.
Power failure.
Contact in-house
electrician.
-
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...