Typical procedure
•
•
Adjust the brightness on the stand by starting with the lowest value and
gradually increasing the brightness level. Please observe the maximum
permissible brightness (see page 26).
•
Adjust the eyepieces (see page 87).
•
Swing the surgical microscope over the surgical field into an ergonomic
position within the working distance.
WARNING
Eye damage!
Looking directly at the light source, e.g. at the microscope lens or optical
fiber, can damage your eyes.
•
Never look directly at the light source!
CAUTION
Retinal damage to the patient’s eye!
Excessive light intensity (brightness setting) or overly long radiation exposure
times may lead to retinal damage in the patient’s eye.
•
Adjust the illumination intensity and the resulting radiation exposure time
by selecting the appropriate illumination setting. You will find the values
recommended by ZEISS in the “Maximum radiation exposure times” table
on page 26.
•
When operating on the eye, use the retina protection filter (blue barrier
filter). It protects the patient’s eye against unnecessary (blue) radiation
and increases the recommended maximum radiation exposure time.
•
If no red reflex is required, swing the retinal protection device into the
beam path.
CAUTION
Variable light intensity due to a faulty light source!
A faulty light source with variable light intensity can damage the patient’s
retina.
•
Check the light source and the brightness adjustment before each use of
the device.
•
Keep a backup lamp or an OR illuminator ready.
•
If the light control fails, switch the device to manual mode (see page 125).
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...