7 Operation
Instructions for Use
OPMI LUMERA 700
7.5 Daily operation
152 / 236
G-30-1673-en - 14.0 - 2018-07-12
5. To save the changes to the surgery profile: tap the
button.
7.4.10 Configuring reset options
83] are basic settings that the device reverts
back to when the XY reset button is pressed on the OPMI or the
device is in the park position.
CAUTION!
Risk of injury caused by lowering of the external focus!
Pressing the XY reset button causes the microscope to focus and
lower, which may injure the patient.
u
Make sure that the working distance between the microscope
and the patient is greater than the travel range of the
microscope.
Prerequisite
þ
The user profile you want to configure is activated.
Action
1. Open the main menu.
2. Tap the [Additional Settings] menu button.
3. Tap the [Reset] submenu button.
ð
The "Reset" menu will appear.
4. Activate the desired functions.
ð
Activated functions are illuminated blue.
5. To save the settings as start values to the user profile: tap the
button.
7.5 Daily operation
During daily operation, settings can be temporarily adjusted, and
certain settings can be saved to the user profile or surgery profile
which is currently active.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...