N HBO 103 and N XBO 75
Start-Up
Microscope Illuminating Systems
Change the Collector
Carl Zeiss
B 40-065 e 08/99
17
3.6
Change the Collector
1 Clamping ring
2 Heat protection filter
3 Collector
4 Groove
5 Collector operating wheel
Fig. 13 Change the collector
•
Loosen clamping ring (13-1)
•
Remove heat protection filter (13-2)
•
Pull out collector operating wheel (13-5)
against resistance until the pin allows access
to groove (13-4)
•
Remove collector from the illuminator
•
Push in new collector until pin of the
pushed-out operating wheel locks into the
groove; the collector can be adjusted via the
operating wheel.
•
Insert heat protection filter;
tighten clamping ring.