NS=çÑ=OO
mä~ëã~=`äÉ~åÉê=^ÇÇ=çå=Ñçê=jìäíápbj
RK=léÉê~íáçå
5.1. Possible Errors
Note
If the plasma cleaner was not in use for a longer time period, the PRS may be oxidized and the
first ignition of the plasma may take several minutes.
If the Plasma Cleaner in the
Plasma Cleaner Control
menu is not connected, check if the power
supply of the plasma cleaner control unit is on. If not, switch power on and press the
Reconnect
button.
The LEDs in the Plasma Cleaner Control
menu reflect the state of the LEDs on the
Plasma cleaner control front panel