Starting the CMM
61211-1270302
MMZ M
6-7
Activating the ram protection (3)
–
Turn the key-operated switch to the «Start» position.
The indicator lamp of the switch then lights up in green. The ram
collision protection is active.
The key-operated switch does not remain in this position. The key-
operated switch returns to the «active» position.
Switching on the drives (4)
–
Press the push button.
The indicator lamp of the switch then lights up. The drives are on.
Selecting the operating mode (5)
The operating mode can also be set later.
–
Turn the key-operated switch to the required position.
MAN
: manual mode - set-up mode. The travel speed is limited in all
three axes.
You must set this operating mode for ...:
–
Joystick-controlled probing
–
Programming measurement runs
–
Testing measurement runs
AUTO
: automatic mode - AUTO operating mode. The CMM travels at
maximum speed in all axes.
You can set this operating mode for ...:
–
Workpiece measurement in a tested measurement run.
Switching the computer on
1
Switch on the computer and peripheral devices such as the printer.
2
Boot the computer; start the operating system.
3
Start the measuring software.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...