Care
9-10
61211-1270302
MMZ M
Adapter plate and adapter plate receptacle
The adapter plate and the adapter plate receptacle must be protected
against soiling.
1. First, clean the adapter plate and the adapter plate receptacle using
Kontakt 60
.
2. Then clean the adapter plate and adapter plate receptacle using
Kontakt WL
.
3. Dry the adapter plate and the adapter plate receptacle in order to
prevent oxidation of the contacts.
NOTICE! Ensure that there are
no cleaning agent residues. The function of the adapter
plate and adapter plate receptacle may be impaired.
NOTE
The adapter plate receptacle must be protected against soiling. Besides,
adapter plates that are not required must be stored in a safe place.
Always leave an adapter plate at the adapter plate receptacle.
Probes and probe holders
Wipe off the probes and probe holders with a dry or damp cloth.
–
Store a probe or a probe holder in a dust-free and clean place, for
example in a cabinet or drawer, whenever it is not in use.
LineScan
The laser beam outlet window on the probe must be free from dirt parti-
cles and smears.
–
Clean the window using a dry, lint-free cloth or a brush.
Changer rack
Vacuum the changer rack and the holders and clean them with a mild
cleaning agent.
–
Remove all cleaning agent residues.
Bellows cover
Vacuum the movable cover and clean it with a damp cloth.
NOTE
The cover must not be removed.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...